Skip to content

Cardiovascular Procedure Kit, P/N 70579-04, St. Louis… recallZ-1313-2008 · Terumo Cardiovascular Systems Corp · initiated Jan 29, 2008

Terminated All recalls posted in March 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Terumo Cardiovascular Systems Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Loose Flash at tip of Weighted Flexible Sucker: A loose strand of stainless steel left on the device during manufacturing by an outside supplier may result in an unretrieved metal fragment remaining in the patient.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Consignees were notified via recall letter, Urgent Medical Device Removal, dated 1/29/08 via fax or email to cease using the weighted sucker in the packs and to use an alternative sucker until replacement product arrives from Terumo.

Recall details

Recall number
Z-1313-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 29, 2008
Posted by the FDA
Mar 28, 2008
Recalling firm
Terumo Cardiovascular Systems Corp
Quantity
252
Product code
OEZ
FDA submissions
K982891
Distribution
Nationwide-USA including states of Missouri, Tennessee and Wisconsin-Custom Kits for 3 hospitals.

Product description

Cardiovascular Procedure Kit, P/N 70579-04, St. Louis University Hospital Pump & Table Pack, Terumo Cardiovascular, Ashland, MA. (the Pump & Table pack contains a weighted flexible sucker, Catalog No. 804113)

Code information (models, lots, serial numbers)

Lot Numbers: GC20, GE24, GF22, GG12, GH17, GL25, GN06, HA22, HD19, HF07, HG18 and HM08.

Other devices under the same FDA submission

4 devices cleared under K982891 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 46869).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page