Skip to content

Biomet Hybrid Glenoid Boss Cutter recallZ-1658-2008 · Biomet, Inc. · initiated Mar 13, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The glenoid implant may not seat fully on the face of the natural glenoid due to inadequate measurement tolerances (tolerance stack-up) in the design of the cutter. Use of the instrument may lead to incomplete seating of the glenoid implant, elevating the theoretical risk of glenoid loosening, which may lead to early revision surgery. The problem may contribute to early device failure and patien

FDA root cause category: Device Design

What the company is doing

Consignees were notified by Urgent Medical Device Removal Notice dated 3/13/08 to discontinue use of the instruments and return to the firm. Operating room managers were sent a separate letter dated 3/13/08 and informed of the possible effects to patients from previous use of this instrument and it was suggested they monitor patients for these effects. If you have any questions contact Mary Hardesty at 1-800-348-9500 or (574) 372-3983.

Recall details

Recall number
Z-1658-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 13, 2008
Posted by the FDA
Sep 16, 2008
Recalling firm
Biomet, Inc.
Quantity
189
Product code
KWT
FDA submissions
K060694
Distribution
Class II Recall - Nationwide Distribution.

Product description

Biomet Hybrid Glenoid Boss Cutter; item 406150. - Surgical instrument used to prepare the natural shoulder glenoid surface for implanting a Hybrid Glenoid component. This instrument is intended to function by drilling a hole into the natural glenoid surface allowing the center boss of glenoid implant to sit within the hole, and the non-articulating surface of the glenoid implant to rest flush onto the natural glenoid surface.

Code information (models, lots, serial numbers)

Lots 44550, 110100, 171760, 211530, 211550, 767940, 820310, 820330, 827250, 853810, 866590, 919780, 919790, 919800, 968270, 980230 and 980250.

Other devices under the same FDA submission

5 devices cleared under K060694 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page