STERRAD NX Sterilizer, Product Code 10033 recallZ-1631-2008 · Advanced Sterilization Products · initiated Feb 25, 2008
Terminated All recalls posted in August 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Sterilization Products or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Oil mist filter fails allowing oil mist to be emitted into the vicinity of the sterilizer, which may result in the release of a mist, "haze," or "smoke."
FDA root cause category: Device Design
What the company is doing
The firm sent an Urgent Device Correction letter dated 4/23/08, notifying customers of a situation that may occur with all models of STERRAD Sterilization Systems (100S, 50, 200, NX and 100NX). As a precaution, personnel should leave the room if mist is observed and avoid returning until the mist has cleared. ASP asks that customers circulate this letter to any STERRAD system users within their facility and also to the Chief Administrative Officer, the Director of Biomedical Engineering, Director of Materials Management and the Chief of Nursing. The letter instructs customers on what to do if they experience a filter failure - a) cancel the cycle if the sterilizer is running [and as always, do not use instruments from an incomplete cycle], b) leave the room, c) discontinue use of the STERRAD Sterilizer until the system is repaired. d) Personnel should avoid working in the room until the mist has cleared. The letter advises that ASP service will replace the sterilizer's filter and increase the frequency of filter changes by adjusting the Planned Maintenance (PM) schedule within the software on their STERRAD 100S and STERRAD NX Sterilizer. Per the notice, for each STERRAD Sterilizer model, ASP will do the following: a) STERRAD 200 Sterilizer - beginning April 28, 2008 through October 2008, ASP will install a new filter and gasket assembly at no charge. b) STERRAD NX Sterilizer - beginning April 28, 2008 through October 2008, ASP will install a new filter assembly at no charge and adjust the PM schedule to 650 cycles or 6 months (whichever occurs first). c) STERRAD 100S Sterilizer - beginning April 28, 2008 through October 2008, ASP will adjust the PM schedule to 750 cycles or 6 months (whichever occurs first). d) STERRAD 50 and 100NX Sterilizers - No adjustments to the oil mist filters or PM schedules are required at this time. PMs will be performed as currently scheduled. Information for MedWatch reporting is included in the letter.
Recall details
- Recall number
- Z-1631-2008
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 25, 2008
- Posted by the FDA
- Aug 8, 2008
- Recalling firm
- Advanced Sterilization Products
- Quantity
- 9,950
- Product code
- MLR
- FDA submissions
- K042116
- Distribution
- Worldwide Distribution -- including USA and countries of France, Germany, Italy, United Kingdom, Switzerland, Benelux, Uruguay, Venezuela, Argentina, Chile, Colombia, Ecuador, Belgium, Puerto Rico, Hong Kong, Indonesia, Mexico, Philippines, Australia, Austria, China, Greece, Ireland, Israel, Japan, Korea, Malaysia, Singapore, Taiwan, Thailand, Turkey, Peru, Canada, Czech Republic, Middle East, Portugal, S. Africa, Spain, Sweden, Hungary, Poland, Brazil, Russia, Egypt, India, and Slovenia.
Product description
STERRAD NX Sterilizer, Product Code 10033 Used to sterilize heat and moisture sensitive reusable medical devices.
Code information (models, lots, serial numbers)
all serial numbers
Other devices under the same FDA submission
1 device cleared under K042116 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- STERRAD NX™ 10033Advanced Sterilization Products Services Inc. · Sterilizer, chemical132 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 47355).
- Z-1628-2008 · TerminatedSTERRAD 100S Sterilization System, Product Code 10101 Used to sterilize heat and moisture sensitive reusable medical devices.
- Z-1629-2008 · TerminatedSTERRAD 50 Sterilization System, Product Code 10050 Used to sterilize heat and moisture sensitive reusable medical devices.
- Z-1632-2008 · TerminatedSTERRAD 100NX Sterilizer, Product Code 10104 Used to sterilize heat and moisture sensitive reusable medical devices.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page