COAT-A-COUNT PSA IRMA kit (IKPS1, 2, 5, X). The device is… recallZ-2090-2008 · Siemens Medical Solutions Diagnostics · initiated May 21, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The field action was initiated after the firm confirmed that the Coat-A-Count PSA IRMA kit (IKPS1, 2, 5, X) exhibits a low bias, which is evident when comparing results with other methods, including the IMMULITE PSA kits. Coat-A-Count PSA IRMA results are, for example, approximately 20% lower when compared to results obtained with the IMMULITE 2000 Third Generation PSA assay (L2KUP).

FDA root cause category: Other

What the company is doing

A Customer Bulletin (Urgent Field Safety Notice) was prepared and sent by fax or mail to all affected Siemens Healthcare Diagnostics Branches on May 21, 2008 both in the United States and elsewhere for communication with the affected customers. This bulletin advises the customer of the problem and the discontinuation of the product. Customers were also instructed to forward this notification to anyone to whom you may have distributed this product and if they have additional questions, please contact your local technical service center @ 1-877-229-3711. A Confirmation (Fax-back) Form is included to ensure that all customers have been notified. The goal is to complete 100% effectiveness by July 31, 2008. If you have questions, contact Siemens Technical Solutions at 1-877-229-3711, option 11, option 6.

Recall details

Recall number
Z-2090-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 21, 2008
Posted by the FDA
Sep 16, 2008
Recalling firm
Siemens Medical Solutions Diagnostics
Quantity
64 kits
Product code
LTJ
FDA submissions
P930027
Distribution
Nationwide to: CA, FL, IL, IN, MD, MI ; Worldwide to: Bolivia, Chile, Egypt, Cyprus, Korea, Philippines, Taiwan, Uruguay

Product description

COAT-A-COUNT PSA IRMA kit (IKPS1, 2, 5, X). The device is intended as an aid in monitoring patients for disease progress or response to therapy.

Code information (models, lots, serial numbers)

LOT NUMBERS: IKPS1 884 - 100 test size, exp. 5/30/2008; IKPS2 864 - 200 test size, exp. 5/30/2008; IKPS5 862 - 500 test size, exp. 5/30/2008; IKPSX 854 - 1000 test size, exp. 5/30/2008

Other devices under the same FDA submission

2 devices cleared under P930027 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.