Permanent Pacing Lead (Catalog #: 03332, 03334, 03337… recallZ-2333-2008 · Oscor, Inc. · initiated Jan 29, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Oscor, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The O-rings are over tolerance making it hard to connect the leads to the pacemaker.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On January 24, 08, Oscor sent out a letter to doctors/ customers with the subject line of "Re: Market Withdrawal of Permanent Pacing Leads due to the lead connector insertion difficulty," stating they experienced some difficulties with the insertion of pacing lead connectors into the pacemaker headers. They asked that customers review the inventory list and remove all products, whose serial numbers match the serial numbers on the list and have not been implanted. Also, customers should call Oscor customer service for a Return Goods Product number and return the product back to the firm for replacement.

Recall details

Recall number
Z-2333-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jan 29, 2008
Posted by the FDA
Sep 24, 2008
Recalling firm
Oscor, Inc.
Product code
DTB
FDA submissions
K010787, K040569, K041809, K872971
Distribution
Nationwide Distribution

Product description

Permanent Pacing Lead (Catalog #: 03332, 03334, 03337, 03107, 03116, 03108, 03109, 05756, 53420, 05765, 03634, 03633, and 05747); Model #: PY244RU, PY252RU, PY258RU, REFINO58RU, REFINO 52RJU, REFINO 44RJU, REFINO48RJU, BIS/IS-15, BIS/BIS-17, VKU-10V, MP52PV, MP40PV, C/IS-10; Manufactured by Oscor. Pacing and sensing of the ventricle and/or atrium of the heart in conjunction with a compatible, implantable pulse generator/pacemaker.

Code information (models, lots, serial numbers)

02-01502, 02-01812, 02-01436, 02-01440, 02-02016, and 02-01945

Other devices under the same FDA submission

2 devices cleared under K010787, K040569, K041809, K872971 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.