Skip to content

MicroScan Synergies Plus Negative Panels, 20 panels per box recallZ-0541-2009 · Siemens Microscan · initiated Jul 31, 2008

Terminated All recalls posted in February 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Microscan or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Falsely lowered imipenem MICs may result, leading to major errors in results.

FDA root cause category: Other

What the company is doing

Consignees were notified on July 31, 2008 by letter (Urgent Device Correction) sent via FedEx overnight. International consignees were notified by Siemens Healthcare Diagnostics associates in affected countries. The letter described the problem and provided users with safety instructions to follow until an update is provided. Contact Siemens Healthcare Diagnostics Inc. at 916-374-3075 for additional assistance.

Recall details

Recall number
Z-0541-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jul 31, 2008
Posted by the FDA
Feb 4, 2009
Recalling firm
Siemens Microscan
Quantity
Total of both Negative and Synergy: 569,375 boxes
Product code
JWY
FDA submissions
K870074
Distribution
Worldwide

Product description

MicroScan Synergies Plus Negative Panels, 20 panels per box Medical device designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility and/or identification to the species level of colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-negative bacteria. Catalog Nos. B1016-200, B1016-201, B1016-202, B1016-203, B1025-103, B1025-104, B1025-105, B1025-106, B1025-107, B1025-108, B1025-109, B1025-112, and B1027-101

Code information (models, lots, serial numbers)

All lots

Other devices under the same FDA submission

1 device cleared under K870074 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page