Skip to content

SOMATOM Emotion 16. Computed tomography X-ray system recallZ-2453-2008 · Siemens Medical Solutions USA, Inc · initiated Aug 5, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential electrical shock hazard. The LCB power Switch (S1) could break causing the protective plastic cap to come off exposing the energized electrical contacts within the switch. The LCB S1 does have an inline fuse which is designed to open in the event of a short circuit condition between the internal switch contacts.

FDA root cause category: Nonconforming Material/Component

What the company is doing

The firm issued and Safety Update in July 2008 informing users that the firm will have representatives visit each account to replace the cap per Update Instructions CT032/08/S.

Recall details

Recall number
Z-2453-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 5, 2008
Posted by the FDA
Sep 21, 2008
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
53 units
Product code
JAK
FDA submissions
K050297
Distribution
Nationwide including states of AK, AL, CO, FL, GA, IA, ID, IL, IN, LA, MD, MN, MS, NC, NE, NY, OH, OK, OR, RI, SD, TX, UT, WA, and WI.

Product description

SOMATOM Emotion 16. Computed tomography X-ray system

Code information (models, lots, serial numbers)

Model Numbers: 10165888, 10165977, and 10046789.

Other devices under the same FDA submission

2 devices cleared under K050297 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page