Skip to content

Plastic Cannula Holdex Tube Holder, Item number: 450216.… recallZ-2198-2008 · Greiner Bio-One North America, Inc. · initiated Aug 5, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Greiner Bio-One North America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

At removal of the tube from the holder the needle may dislodge and become stuck in the tube stopper with the blunt end of the needle facing out.

FDA root cause category: Process design

What the company is doing

Consignees were notified by letter on 08/06/2008. Letters to Distributors instructed them to stop distribution of the affected product immediately and to hold any product of the lots they may have pending returned to Greiner bio-one. The were instructed to provide Greiner bio-one with a list of their customers who may have received the affected lot so that they can contact them. A Product Disposition form was included to be completed and returned to Greiner bio-one. Contact Greiner Bio-One North America, Inc. at 1-888-286-3883 for assistance.

Recall details

Recall number
Z-2198-2008
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 5, 2008
Posted by the FDA
Sep 30, 2008
Recalling firm
Greiner Bio-One North America, Inc.
Quantity
150,000 pcs
Product code
FMI
FDA submissions
K980768
Distribution
AL, FL, IN, KS, KY, MO, MN, NC, OH, PA, TN, TX, UT, VA, WI and WV.

Product description

Plastic Cannula Holdex Tube Holder, Item number: 450216. It is intended for use in routine specimen withdrawal from specimen collection bags or bottles through needleless cannula ports.

Code information (models, lots, serial numbers)

Lot#: A030804

Other devices under the same FDA submission

1 device cleared under K980768 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.