Roche ISE Internal Standard Gen.2 for use on Roche cobas c… recallZ-0542-2009 · Roche Diagnostics Corp. · initiated Sep 15, 2008

Terminated All recalls posted in January 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Roche Diagnostics Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Test results, low: Sodium, potassium and chloride results on the cobas 6000 ISE unit may exhibit a downward drift in controls and patient samples within the recommended calibration interval of 24 hours. For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.

FDA root cause category: Package design/selection

What the company is doing

Consignees were notified by letter dated 9/15/08, which provided a work-around for them to use until the problem can be permanently resolved. The letter was entitled "URGENT MEDICAL DEVICE CORRECTION." For further information, please contact Roche Diagnostics Corporation by telephone at 800-428-2336.

Recall details

Recall number
Z-0542-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 15, 2008
Posted by the FDA
Jan 6, 2009
Recalling firm
Roche Diagnostics Corp.
Quantity
7909
Product code
JIX
FDA submissions
K060373
Distribution
Nationwide Distribution.

Product description

Roche ISE Internal Standard Gen.2 for use on Roche cobas c systems, Roche Diagnostics, Inc., Indpls., IN; Catalog no. 04522320190. Product is an in vitro diagnostic.

Code information (models, lots, serial numbers)

Lot numbers: 60258801, 60258601, 60145701, 60145601, 69921901, 69921801, 69786201, 69922001, 69786101, 69676101, 69509701, 69431901, 69357601, 69342001, 69246501, 68966601 and 68871601.

Other devices under the same FDA submission

11 devices cleared under K060373 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.