VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator)… recallZ-0175-2009 · ConMed Corporation · initiated Sep 8, 2008
Terminated All recalls posted in October 2008
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ConMed Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm received several complaints sighting detachment of the balloon at the distal end of the shaft. It was also noted that the incidence of the forward cup slipping off the shaft and being retained in the vaginal canal had increased.
FDA root cause category: Process change control
What the company is doing
The domestic recall notices were mailed on 9/08/08 by USP Priority Mail pouches. The international notices were sent via FedEx International Priority on 9/08/08. The letter described the problem, the health risk and requested that customers inventory the product and return them back to the firm. Customers are requested to fax the reply form to the firm. Contact ConMed Corporation at 1-800-765-8375, ext 3237 for assistance.
Recall details
- Recall number
- Z-0175-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Sep 8, 2008
- Posted by the FDA
- Oct 28, 2008
- Recalling firm
- ConMed Corporation
- Quantity
- 32,453 (total devices)
- Product code
- LFK
- FDA submissions
- K071907
- Distribution
- Nationwide, Australia, Belgium, Brazil, Chile, Ecuador, France, and Switzerland
Product description
VCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) STANDARD, Uterine Manipulator, REF 60-6085-000. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502. The ConMed VCARE Retractor/Elevator is indicated for manipulation of the uterus and injection of fluids or gases during laparoscopic procedures such as Laparoscopic Assisted Vaginal Hysterectomy (LAVH), Total Laparoscopic Hysterectomy (TLH), minilap, laparoscopic tubal occlusion, or diagnostic laparoscopy and also maintains pneumoperitoneum by sealing the vagina once colpotomy is performed.
Code information (models, lots, serial numbers)
Lots 0710121x through and including 0804111x [yymmddx where x = shift code 1,2, or 3] . All lots of the recalled devices were manufactured between and including October 12, 2007 and April 11, 2008.
Other devices under the same FDA submission
3 devices cleared under K071907 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- VCARE MEDIUM 60-6085-201Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE9 reports
- VCARE SMALL 60-6085-200Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE6 reports
- VCARE LARGE 60-6085-202Conmed Corporation · CANNULA, MANIPULATOR/INJECTOR, UTERINE1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 49406).
- Z-0176-2009 · TerminatedVCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) SMALL, Uterine Manipulator, REF 60-6085-001. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502. The ConMed VCARE Re…
- Z-0177-2009 · TerminatedVCARE (Vaginal-Cervical-Ahluwalia's-Retractor-Elevator) LARGE, Uterine Manipulator, REF 60-6085-002. Made in USA, CONMED CORPORATION, UTICA, NEW YORK USA 13502. The ConMed VCARE …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page