Why it was recalled
Sporadic errors in various device functions, including low readings of the single gas flows for oxygen, nitrous oxide, or air delivery, or mechanical ventilation failure. Alarms and error messages function as designed.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Draeger Medical sent a Medical Device Recall letter to their customers via certified mail on 8/22/08 informing them of the problem. Until the exchange of affected printed circuit boards has been implemented, the firm is recommending that users closely monitor all device displays and error messages and if applicable act in accordance with the advisory statements and remedial information provided in the Instructions for Use as well as making provisions for alternative patient care measures.
Recall details
- Recall number
- Z-0087-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Aug 22, 2008
- Posted by the FDA
- Oct 8, 2008
- Recalling firm
- Draeger Medical, Inc.
- Quantity
- 214 units
- Product code
- BSZ
- FDA submissions
- K042607
- Distribution
- Nationwide Distribution.
Product description
Draeger Medical Apollo Anesthesia Machine, catalog number 8606500
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
Catalog number 8606500. Serial numbers ARZC-0142, ARZC-0143, ARZC-0144, ARZC-0145, ARZC-0146, ARZC-0147, ARZC-0148, ARZC-0149, ARZC-0150, ARZC-0151, ARZC-0152, ARZC-0153, ARZC-0172, ARZC-0173, ARZC-0174, ARZC-0175, ARZC-0176, ARZC-0177, ARZC-0192, ARZC-0193,ARZC-0196, ARZC-0197, ARZC-0198, ARZC-0199, ARZC-0200, ARZD-0001, ARZD-0002, ARZD-0003, ARZD-0010, ARZD-0011, ARZD-0012, ARZD-0013, ARZD-0014, ARZD-0015, ARZD-0016, ARZD-0017, ARZD-0018, ARZD-0019, ARZD-0020, ARZD-0021, ARZD-0022, ARZD-0023, ARZD-0024, ARZD-0025, ARZD-0026, ARZD-0027, ARZD-0028, ARZD-0029, ARZD-0030, ARZD-0031, ARZD-0032, ARZD-0033, ARZD-0034, ARZD-0035, ARZD-0036, ARZD-0037, ARZD-0038, ARZD-0039, ARZD-0040, ARZD-0041, ARZD-0042, ARZD-0043, ARZD-0044, ARZD-0058, ARZD-0059, ARZD-0060, ARZD-0061, ARZD-0062, ARZD-0063, ARZD-0064, ARZD-0065, ARZD-0066, ARZD-0067, ARZD-0068, ARZD-0069, ARZD-0070, ARZD-0071, ARZD-0072, ARZD-0073, ARZD-0074, ARZD-0075, ARZD-0076, ARZD-0077, ARZD-0078, ARZD-0079, ARZD-0080, ARZD-0081, ARZD-0082, ARZD-0083, ARZD-0084, ARZD-0085, ARZD-0086, ARZD-0087, ARZD-0117, ARZD-0118, ARZD-0119, ARZD-0120, ARZD-0121, ARZD-0122, ARZD-0125, ARZD-0126, ARZD-0127, ARZD-0128, ARZD-0129, ARZD-0130, ARZD-0131, ARZD-0132, ARZD-0133, ARZD-0134, ARZD-0135, ARZD-0136, ARZD-0183, ARZD-0184, ARZD-0185, ARZD-0186, ARZD-0187, ARZD-0188, ARZD-0189, ARZD-0190, ARZD-0191, ARZD-0192, ARZD-0193, ARZD-0194, ARZD-0195, ARZD-0196, ARZD-0197, ARZD-0198, ARZD-0199, ARZD-0200, ARZD-0201, ARZD-0202, ARZD-0203, ARZD-0255, ARZD-0256, ARZD-0257, ARZD-0258, ARZD-0259, ARZD-0260, ARZD-0261, ARZD-0262, ARZD-0263, ARZD-0264, ARZD-0265, ARZD-0266, ARZD-0267, ARZD-0268, ARZD-0269, ARZD-0270, ARZD-0271, ARZE-0001, ARZE-0002, ARZE-0003, ARZE-0004, ARZE-0005, ARZE-0008, ARZE-0009, ARZE-0010, ARZE-0023, ARZE-0024, ARZE-0025, ARZE-0026, ARZE-0027, ARZE-0028, ARZE-0029, ARZE-0030, ARZE-0059, ARZE-0060, ARZE-0061, ARZE-0062, ARZE-0063, ARZE-0064, ARZE-0065, ARZE-0115, ARZE-0116, ARZE-0173, ARZE-0174, ARZE-0175, ARZE-0176, ARZE-0189, ARZE-0190, ARZE-0191, ARZE-0192, ARZE-0193, ARZE-0194, ARZE-0195, ARZE-0196, ARZE-0197, ARZE-0198, ARZE-0199, ARZE-0200, ARZE-0201, ARZE-0202, ARZE-0203, ARZE-0204, ARZE-0214, ARZE-0215, ARZE-0216, ARZF-0011, ARZF-0012, ARZF-0013, ARZF-0014, ARZF-0015, ARZF-0016, ARZF-0017, ARZF-0018, ARZF-0019, ARZF-0020, ARZF-0021, ARZF-0022, ARZF-0023, ARZF-0024, ARZF-0025, ARZF-0026, ARZF-0027, ARZF-0028, ARZF-0029, ARZF-0030, ARZF-0031, ARZF-0084, ARZF-0087, ARZF-0088, ARZF-0089, ARZF-0090, ARZF-0091, ARZF-0092, ARZF-0093, ARZF-0094, ARZF-0095, ARZF-0096, ARZF-0097, ARZF-0098, ARZF-0099, ARZF-0100, ARZF-0101, ARZF-0102, ARZF-0103, ARZF-0104, ARZF-0105, ARZF-0106, ARZF-0107, ARZF-0108, ARZF-0109, ARZF-0110, and ARZF-0111.
Other devices under the same FDA submission
1 device cleared under K042607 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Apollo 8606500Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia90 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page