Weck Hemoclip Traditional Ligating Clips, Rx Only, Sterile… recallZ-0156-2009 · Telefelx Medical · initiated Aug 22, 2008

Terminated All recalls posted in September 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Telefelx Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A hole in the sterile unit blister pack was detected that would compromise sterility.

FDA root cause category: Packaging process control

What the company is doing

Consignees were notified via UPS next day on/about August 22, 2008. They were instructed to cease use and distribution and quarantine all affected product immediately. If they have any of the affected product they were instructed to return all it to Teleflex Medical via provided UPS-Ground. An Acknowledgement form was provided to be completed and returned via fax to Teleflex Medical to confirm receipt of the notification and amount of product on hand for return. A second mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks from the second mailing. Consignees also may be contacted by phone, e-mail, or fax. A record of these contacts will be maintained. Upon return to Teleflex Medical, product will be quarantined, assessed for recall effectiveness, and either reworked or destroyed. Contact Telefex Medical at 1-800-334-9751 ext. 8066 for assistance.

Recall details

Recall number
Z-0156-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 22, 2008
Posted by the FDA
Sep 16, 2008
Recalling firm
Telefelx Medical
Quantity
786036 (total pieces)
Product code
FZP
FDA submissions
K021808
Distribution
Nationwide

Product description

Weck Hemoclip Traditional Ligating Clips, Rx Only, Sterile, Teleflex Medical, RTP, NC 27709 USA. The use of the reusable Teleflex Medical metal ligating clips is transient, invasive and intended for the application of hemostatic clips during general open surgery. Ligation clips are non-absorbable to thermostatically restrict flow of fluids within vessels. Clips are removed from the cartridge and applied in such a manner as to engulf vessels. Teleflex Medical metal ligating clips are intended for use in procedures involving vessels or anatomic structures for which the surgeon determines ligating clips are the best choice. Surgeons should select the size type, and material of the clip based upon their experience, judgment and needs.

Code information (models, lots, serial numbers)

Show the full code information (9k characters)

Product manufactured from January 2006 through July 2008; Product Description: HEMOCLIP MEDIUM 250/BOX, Product Code: 523100, Lot number: 01C0800066, 01D0800156, 01D0800324, 01E0800203, 01E0800410, 01K0700113, 01K0700198, 01K0700298, 01L0700009, 01M0700131, T1202225, T1202961, T1209529, T1211719, T1213975, T1220115, T1226528, T1228532, T1229942, T1233481, T1235583, T1236588, T1239523, T1242799, T1248517, T1249582, T1250496, T1252394, T1255378, T1256389, T1258311, T1261008, T1262892, T1264925, T1270366, T1272027, T1273754, T1275784, T1277275, and T1280263; Product Description: HEMOCLIPS SMALL 300/BOX, Product Code: 523135, Lot number: 01A0800221, 01B0800284, 01E0800020, 01E0800051, 01E0800205, 01K0700299, 01L0700097, 01M0700054, T1202223, T1207485, T1210251, T1218339, T1224895, T1229939, T1231420, T1235581, T1239179, T1241604, T1243428, T1247446, T1248346, T1256385, T1258309, T1260080, T1261781, T1270990, T1272023, and T1276617; Product Description: HEMOCLIP MED-LGE 200/BOX, Product Code: 523160, Lot number: 01K0700199, 01L0700098, T1201401, T1202195, T1210839, T1213979, T1229935, T1237417, T1240028, T1242797, T1260076, and T1271937; Product Description: HEMOCLIP LARGE 150/BOX, Product Code: 523170, Lot number: 01A0800325, 01D0800157, 01E0800018, 01L0700340, T1198153, T1201448, T1203611, T1210220, T1214463, T1218859, T1228527, T1237522, T1238441, T1242798, T1246173, T1248344, T1253536, T1254318, T1255373, T1258761, T1262886, T1264105, T1276611, and T1280177; Product Description: HEMOCLIP TEN PAK MEDIUM 200/BOX, Product Code: 523300, Lot number: 01B0800289, 01D0800158, 01F0800071, 01K0700200, 01K0700300, 01L0700010, 01M0700055, T1195698, T1203615, T1208575, T1213976, T1216526, T1227714, T1228534, T1231501, T1233521, T1234311, T1252395, T1254402, T1260323, T1262893, T1266943, T1270995 and T1276883; Product Description: HEMOCLIP TEN PAK SMALL 240/BOX, Product Code: 523335, Lot number: 01A0800327, 01C0800296, 01D0800159, T1203612, T1213977, T1221073, T1225877, T1230388, T1237120, T1239180, T1241605, T1251449, T1253540, T1260321, T1261792, T1273752, and T1277291; Product Description: HEMOCLIP TEN PAK MED-LG 160/BX, Product Code: 523360, Lot number: 01A0800328, 01C0800244, T1195769, T1196987, T1199780, T1208571, T1218337, T1229936, T1237418, T1247368, T1255369, T1264918, T1271938, and T1273748; Product Description: HEMOCLIP TEN PAK LG 120/BOX, Product Code: 523370, Lot number: 01C0800245, T1202199, T1209526, T1211715, T1227696, T1234309, T1238442, T1239176, T1243426, T1245243, T1248508, T1250484, T1254319, T1267799, T1268809, T1275639, T1276612, T1278104, and T1279053; Product Description: HEMOCLIP S/S MEDIUM 250/BOX, Product Code: 523400, Lot number: 01A0800329, 01D0800086, 01E0800409, T1226527, T1226762, T1229941, T1232352, T1237527, T1238445, T1248515, T1249520, T1250494, T1251452, T1261740, T1262890, T1264923, T1266142, T1270993, T1272025, T1275783, T1278107, and T1279058; Product Description: HEMOCLIP S/S SMALL 300/BOX, Product Code; 523435, Lot number: 01E0800144, 01L0700182, 01M0700132, T1200641, T1232348, T1239177, T1247445, T1255376, T1257214, T1262887, and T1276616; Product Description: HEMOCLIP S/S MED/LG 200/BOX, Product Code: 523460, Lot number: 01B0800076, 01D0800234, 01L0700258, T1201481, T1203609, T1206790, T1212659, T1219274, T1224891, T1234306, T1239175, T1245241, T1252391, T1255370, T1271939, and T1277285; Product Description: HEMOCLIP S/S LARGE 150/BOX, Product Code: 523470, Lot number: 01B0800163, 01C0800064, 01L0700171, 01M0700133, T1214678, T1226761, T1229937, T1237523, T1240293, T1255374, T1257213, T1260078, T1267800, T1276613, T1277287, and T1280178; Product Description: HEMOCLIP S/S MEDIUM 200/BOX, Product Code: 523600, Lot number: 01B0800389, 01D0800160, 01E0800021, 01L0700099, 01M0700134, T1211718, T1216521, T1227711, T1230389, T1238446, T1242668, T1247371, T1251453, T1256388, T1262891, T1264924, T1266942, T1270994, T1272879, T1277274, and T1280262; Product Description: HEMOCLIP S/S SMALL 240/BOX, Product Code: 523635, Lot number: 01A0800050, 01A0800169, 01C0800143, 01C0800246, 01C0800295, 01D0800236, T1201422, T1202222, T1206018, T1209527, T1211716, T1213978, T1221290, T1224894, T1227708, T1230387, T1233478, T1237119, T1239178, T1240295, T1253538, T1259106, T1262888, T1266119, T1268812, T1270989, T1273751, T1278105, and T1279054; Product Description: HEMOCLIP S/S MEDIUM-LARGE 160/BOX, Product Code: 523660, Lot number: 01A0800037, T1202196, T1256382, T1262885, and T1271940; Product Description: HEMOCLIP S/S LARGE 120/BOX, Product Code: 523670, Lot number: 01A0800051, 01C0800065, T1228528, T1237524, T1245245, T1248345, T1266118, and T1277288; Product Description: HEMOCLIP TITANIUM MEDIUM 250/BOX, Product Code: 523700, Lot number: 01A0800052, 01A0800170, 01A0800331, 01A0800426, 01A0800427, 01B0800390, 01D0800087, 01D0800088, 01D0800283, 01D0800325, 01E0800412, 01K0700286, 01L0700100, 01L0700342, T1202226, T1203616, T1214464, T1222328, T1223063, T1224897, T1226763, T1230390, T1231505, T1232353, T1233482, T1234312, T1235584, T1236584, T1238447, T1244668, T1248519, T1249585, T1252396, T1253543, T1255379, T1255380, T1257215, T1258314, T1259111, T1260324, T1261794, T1262894, T1264108, T1266144, T1266944, T1268384, T1270368, T1270996, T1272028, T1272029, T1273755, T1275785, and T1279059; Product Description: HEMOCLIP TITANIUM SMALL 300/BOX, Product Code: 523735, Lot number: 01B0800315, 01C0800145, 01C0800297, 01D0800395, 01E0800146, 01E0800204, T1192807, T1195697, T1202224, T1216520, T1221291, T1222327, T1223064, T1226526, T1228531, T1229940, T1232349, T1236587, T1237526, T1239521, T1241480, T1243837, T1245247, T1249518, T1251450, T1252393, T1253541, T1258215, T1259108, T1261004, T1262889, T1266941, T1268815, T1270364, T1270991, T1272877, T1273753, T1275781, T1276620, T1278106, and T1279056; Product Description: HEMOCLIP TITANIUM MEDIUM-LARGE 200/BOX, Product Code: 523760, Lot number: 01A0800222, 01B0800077, 01D0800008, 01D0800164, 01E0800143, T1202197, T1208572, T1210838, T1234307, T1237419, T1247369, T1252392, T1256383, T1257719, T1258560, T1259103, T1261790, T1266799, T1272021, T1273749, and T1277286; Product Description: HEMOCLIP TITANIUM LARGE 150/BOX, Product Code: 523770, Lot number: 01A0800054, 01D0800235, 01D0800322, 01L0700011, 01L0700260, T1202200, T1215225, T1220112, T1224663, T1227699, T1228529, T1237525, T1238443, T1244223, T1247370 , T1248510 , T1250486 , T1254320, T1259104, T1264919, T1272875, T1276614, and T1280179; Product Description: HEMOCLIP TITANIUM MEDIUM 200/BOX, Product Code: 523800, Lot number 1A0800223, 01A0800428, 01B0800165, 01C0800247, 01C0800299, 01D0800089, 01D0800288, 01D0800337, 01E0800023, 01K0700301, 01L0700012, 01L0700178, 01L0700261, 01L0700343, 01M0700136, T1192808, T1202227, T1206021, T1207491, T1213973, T1222336, T1223062, T1224085, T1226764, T1228535, T1229944, T1231510, T1232354, T1233483, T1234314, T1235585, T1238448, T1239525, T1241011, T1241602, T1242669, T1243437, T1244224, T1244667, T1247447, T1248348, T1249587, T1250498, T1252397, T1252398, T1253544, T1254404, T1256391, T1258216, T1259112, T1260326, T1260617, T1261795, T1261796 T1264109, T1264926, T1266145, T1266946, T1268385, T1270369, T1270997, T1272880, T1273756, T1275787, T1276885, T1278109, and T1279060; Product Description: HEMOCLIP TITANIUM SMALL 240/BOX, Product Code: 523835, Lot number: 01A0800171, 01A0800224, 01C0800146, 01C0800248, 01C0800298, 01D0800146, 01D0800161, 01D0800323, 01E0800290, 01F0800084, 01K0700114, 01K0700202, 01K0700302, 01L0700161, 01L0700262, 01L0700344, 01M0700137, T1195699, T1208574, T1213974, T1217528, T1224083, T1227710, T1232350, T1233479, T1235582, T1241601, T1246175, T1248347, T1249519, T1251451, T1256386, T1258310, T1260322, T1261006, T1264107, T1264922, T1266141, T1268381, T1268816, T1270365, T1270992, T1272878, T1275782, T1276881, and T1277293; Product Description: HEMOCLIP TITANIUM MEDIUM-LARGE 160/BOX, Product Code: 523860, Lot number: 01K0700203, 01L0700013, 01L0700183, 01M0700138, T1195700, T1201399, T1202198, T1202860, T1203610, T1206016, T1213980, T1218338, T1226600, T1227690, T1230384, T1234308, T1235439, T1237521, T1238396, T1240292, T1244666, T1245242, T1248343, T1253535 , T1255372, T1256384, T1260077, T1261791, T1262882, T1264104, T1266117, T1266800, T1272022, T1273750, T1278103, and T1279050; Product Description: HEMOCLIP TITANIUM LARGE 120/BOX, Product Code: 523870, Lot number: 01A0800055, 01B0800388, 01C0800294, 01D0800165, 01E0800019, 01E0800206, 01K0700303, 01L0700014, 01L0700184, T1192809, T1202221, T1212660, T1220113, T1225876, T1227701, T1230386, T1236585, T1238444, T1244676, T1246174, T1248512, T1253537, T1254401, T1255375, T1258762, T1260079, T1261782, T1264920, T1268810, T1272876, T1276615, and T1280180.

Other devices under the same FDA submission

1 device cleared under K021808 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 49452).