Skip to content

Novel PEEK VBR XS 16-22 MM, S-5, 26MM, Part Numbers :… recallZ-0508-2009 · Alphatec Spine, Inc. · initiated Aug 16, 2007

Terminated All recalls posted in December 2008

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Alphatec Spine, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Firm confirmed that two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. They were incorrectly laser marked with the wrong part numbers and sizes. Two lots were mixed from the following lot numbers: Part Number 64733-120 Lot Number 611884, and Part Number 64753-126 and Lot Number 611881.

FDA root cause category: Process control

What the company is doing

The recall was initiated on August 16, 2007 with Alphatec Spine contacting each of the four direct distributors by telephone. This was followed with a Product Recall Notice, dated August 27, 2007, that was mailed to all affected customers. The distributors were informed that the firm was conducting a voluntary recall because two lots were mixed during manufacturing and were incorrectly labeled with the wrong part number & lot number. The firm asked the customers to return the product, and were informed that replacement product will be sent back.

Recall details

Recall number
Z-0508-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Aug 16, 2007
Posted by the FDA
Dec 31, 2008
Recalling firm
Alphatec Spine, Inc.
Quantity
4 units
Product code
MQP
FDA submissions
K050553
Distribution
Nationwide Distribution --- including states of AZ, FL, NY & TX.

Product description

Novel PEEK VBR XS 16-22 MM, S-5, 26MM, Part Numbers : 64753-126. Indicated for use in the thoracolumbar spine (T1 to L5) for partial or total replacement of a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). Intended for use with supplemental spinal fixation systems.

Code information (models, lots, serial numbers)

Lot Number: 611881

Other devices under the same FDA submission

1 device cleared under K050553 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page