Skip to content

DLP(R) Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N:… recallZ-0609-2009 · Medtronic Cardiovascular Revascularization & Surgical Therap · initiated Sep 29, 2008

Terminated All recalls posted in January 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medtronic Cardiovascular Revascularization & Surgical Therap or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Separates: Six complaints were received from two locations. In all cases the sump tip assembly separated from the tubing during device removal and had to be retrieved from the patient. There were no adverse events affecting these patients.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Consignees were sent a "Medtronic Urgent Medical Device Recall Notice" dated October 1, 2008. The letter described the product and issue. It also requested its customers to return the product and to complete, sign and fax the Recall Certificate "Medtronic Recall Certificate (FCA 0902)". For further information, please contact Medtronic Cardiovascular Revascularization & Surgical Therap by telephone at 763-391-9000.

Recall details

Recall number
Z-0609-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 29, 2008
Posted by the FDA
Jan 26, 2009
Recalling firm
Medtronic Cardiovascular Revascularization & Surgical Therap
Quantity
380 US ; 180 OUS
Product code
DWF
FDA submissions
K850385
Distribution
Worldwide Distribution --- including states of NC, PA, IA, WI, IN, VA, IL, KY, and TX, and countries of Australia, France, Hong Kong, Netherlands, and Poland.

Product description

DLP(R) Pericardial/Intracardiac Sump Cannulae, 20 Fr., P/N: 12112. This product is intended for draining the pericardial sac or the cardiac chambers.

Code information (models, lots, serial numbers)

Lot number 2006100305

Other devices under the same FDA submission

3 devices cleared under K850385 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page