Medtronic Snap Shunt Ventricular Catheter, Standard, with… recallZ-1144-2009 · Medtronic Neurosurgery · initiated Feb 12, 2009
Terminated All recalls posted in March 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Neurosurgery or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The ventricular catheters may become detached from the Snap Base assembly after implantation and this condition may increase the need for emergency revision surgery.
FDA root cause category: Device Design
What the company is doing
An Urgent - Medical Device Recall notification was sent on February 13,2009 to the attention of risk managers, advising them to cease use of the Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782 ( all versions and all lot numbers), the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27802 ( all versions and all lot numbers), and the Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27708 ( all version and all lot numbers). Each customer is to cease the use of the listed product and any unused unit needs to be returned to Medtronic. Further, each account is required to account for product used/implanted and to advise surgeons of the recall information. Each account will recieved an individualized Customer Product Reconciliation Form provided by their Medtronic Sales Representative for determining the status of any product shipped to them.
Recall details
- Recall number
- Z-1144-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 12, 2009
- Posted by the FDA
- Mar 27, 2009
- Recalling firm
- Medtronic Neurosurgery
- Quantity
- 3048 of all units
- Product code
- JXG
- FDA submissions
- K792007, K874498, K951258, K983331
- Distribution
- Worldwide Distribution -- USA, Ireland, and Australia.
Product description
Medtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 3 CM, Ref 27802A. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The sealed pouch is then placed within a chipboard carton along with product IFU and patient wallet cards. The product is ETO sterilized.
Code information (models, lots, serial numbers)
All Lots
Other devices under the same FDA submission
28 devices cleared under K792007, K874498, K951258, K983331 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- BioGlide® 91503Medtronic Ps Medical, Inc. · Shunt, central nervous system and components23 reports
- NA 41101Medtronic Ps Medical, Inc. · Shunt, central nervous system and components22 reports
- BioGlide® 91503Medtronic Ps Medical, Inc. · Shunt, central nervous system and components17 reports
- N/A 41101Medtronic Ps Medical, Inc. · Shunt, central nervous system and components14 reports
- N/A 22011 MMedtronic Ps Medical, Inc. · Shunt, central nervous system and components10 reports
- BioGlide® 93522Medtronic Ps Medical, Inc. · Shunt, central nervous system and components8 reports
- NA 22011 MMedtronic Ps Medical, Inc. · Shunt, central nervous system and components4 reports
- Delta®, BioGlide® 96823Medtronic Ps Medical, Inc. · Shunt, central nervous system and components4 reports
- NA 22011 LMedtronic Ps Medical, Inc. · Shunt, central nervous system and components2 reports
- NA 41406Medtronic Ps Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR2 reports
- NA 41407Medtronic Ps Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVENTRICULAR2 reports
- Innervision 27240Medtronic Ps Medical, Inc. · Shunt, central nervous system and components2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51287).
- Z-1124-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782L. Sizes: 3 CM in length, REF (240) 27782L. Each catheter is provided in a sealed Tyvek plas…
- Z-1125-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782K. Sizes:5.5CM in length, REF (240) 27782K. Each catheter is provided in a sealed Tyvek plas…
- Z-1126-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782J. Sizes: 4.5 CM in length, REF (240) 27782J. Each catheter is provided in a sealed Tyvek pl…
- Z-1127-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782H. Sizes: 11 CM in length, REF (240) 27782H. Each catheter is provided in a sealed Tyvek pla…
- Z-1128-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782G. Sizes 10cm in length, REF (240) 27782G. Each catheter is provided in a sealed Tyvek plast…
- Z-1129-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782F. Sizes: 9 CM in length, REF (240) 27782F. Each catheter is provided in a sealed Tyvek plas…
- Z-1130-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782E. Sizes: 8 CM in length, REF (240) 27782E. Each catheter is provided in a sealed Tyvek plas…
- Z-1131-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782D. Sizes: 7 CM in length, REF (240) 27782D. Each catheter is provided in a sealed Tyvek plas…
- Z-1132-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782C. Sizes: 6 CM in length, REF (240) 27782C. Each catheter is provided in a sealed Tyvek plas…
- Z-1133-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782B. Sizes: 5 CM in length, REF (240) 27782B. Each catheter is provided in a sealed Tyvek plas…
- Z-1134-2009 · TerminatedMedtronic Innervision Snap Shunt Ventricular Catheter, BioGlide, Catalog Number 27782A. Sizes: 4 CM in length, REF (240) 27782 a. Each catheter is provided in a sealed Tyvek pla…
- Z-1135-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 4 CM, Ref 27708A. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1136-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 5 CM, Ref 27708B. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1137-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 6 CM, Ref 27708C. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1138-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 7 CM, Ref 27708D. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1139-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 8 CM, Ref 27708E. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1140-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 3 CM, Ref 27708F. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1141-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 9CM, Ref 27708G. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The s…
- Z-1142-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 10 CM, Ref 27708H. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The…
- Z-1143-2009 · TerminatedSnap Shunt Ventricular Catheter, Standard, with BioGlide, 11 CM, Ref 27708J Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar pouch. The …
- Z-1145-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 4 CM, Ref 27802B. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1146-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 5 CM, Ref 27802C. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1147-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 6 CM, Ref 27802D. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1148-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 7 CM, Ref 27802E. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1149-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 8 CM, Ref 27802F. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1150-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 9 CM, Ref 27802G. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar po…
- Z-1151-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 10 CM, Ref 27802H. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar p…
- Z-1152-2009 · TerminatedMedtronic Snap Shunt Ventricular Catheter, Standard, with BioGlide, 11 CM, Ref 27802J. Each catheter is provided in a sealed Tyvek plastic tray placed within a sealed Tyvek/Mylar p…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page