STERRAD 50 Sterilization System Product Code 10050 recallZ-1419-2009 · Advanced Sterilization Products · initiated Mar 9, 2009
Terminated All recalls posted in June 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Advanced Sterilization Products or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ASP has identified a secondary, lower-occurring cause of oil mist within certain STERRAD NX Systems and STERRAD 50 Systems due to potentially defective oil fill plugs.
FDA root cause category: Component design/selection
What the company is doing
Firm mailed out Urgent Device Correction letters dated March 9, 2009, advising customers/users that on April 23, 2008, Advanced Sterilization Products (ASP) initiated a device correction notification related to certain models of STERRAD Sterilization Systems that have the potential to emit oil mist due to a premature mechanical failure in the oil mist filter and/or filter assemblies. All affected STERRAD Systems have been corrected since the notification was issued. Recently, a secondary, lower-occurring cause of oil mist has been identified within the STERRAD NX System and STERRAD 50 System oil fill plugs. STERRAD NX System and/or STERRAD 50 System may potentially have defective oil fill plugs. Less than 1% of the installed STERRAD NX Systems have been affected. ASP has received 18 complaints in 10 months related to this failure for the STERRAD NX System. No complaints related to this issue for the STERRAD 50 System have been received. As of December 22, 2008, ASP has been actively replacing the plugs in these systems with stainless steel plugs during service calls and will continue to replace the oil fill plugs until all affected units are corrected. The letter advises that there is a risk of symptoms, such as nausea, shortness of breath, dry throat, burning, watery and blurry eyes, and headache occurring if customers inhale oil mist. If you see mist, haze or smoke in the room or detect an oil odor, personnel should complete the following: - Cancel the cycle if the sterilizer is running and as always, do not use instruments from an incomplete cycle - Leave the room - Avoid working in the room until the mist has cleared - Discontinue use of the STERRAD System until the system is repaired. Any additional questions can be directed to the Clinical and Technical Support staff at 1-888-4613.
Recall details
- Recall number
- Z-1419-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 9, 2009
- Posted by the FDA
- Jun 15, 2009
- Recalling firm
- Advanced Sterilization Products
- Quantity
- 2,209 units installed in US; 2013 units OUS
- Product code
- FLF
- FDA submissions
- K023290, K981625
- Distribution
- Worldwide Distribution -- USA, Germany, Italy, France, UK, Switzerland, Benelex, Uruguay, Venezuela, Argentina, Chile, Columbia, Ecuador, Belgium, Puerto Rico, Hong Kong, Indonesia, Mexico, Philippines, Austria, China, Greece, Ireland, Israel, Japan, Korea, Malaysia, Singapore, Thailand, Taiwan, Turkey, Middle East, Peru, Canada, Czech Republic, Portugal, South Africa, Spain, Sweden, Hungary, Poland, Brazil, Russia, Egypt, India, and Slovenia.
Product description
STERRAD 50 Sterilization System Product Code 10050 The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive reusable medical devices.
Code information (models, lots, serial numbers)
All serial numbers
Other devices under the same FDA submission
2 devices cleared under K023290, K981625 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Sterrad 10101Advanced Sterilization Products Services Inc. · Sterilizer, ethylene-oxide gas258 reports
- STERRAD™ 10113Advanced Sterilization Products Services Inc. · Sterilizer, chemical2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51413).
- Z-1420-2009 · TerminatedSTERRAD NX Sterilization System Product Code 10033 The STERRAD Sterilization System is a low-temperature general-purpose sterilizer used to sterilize heat and moisture sensitive r…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page