Phoenix Cannulated Recon Screw recallZ-2115-2009 · Ebi, L.P. · initiated Mar 27, 2009
Terminated All recalls posted in September 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ebi, L.P. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain lots of raw material contain inclusions that could potentially reduce the strength of some devices.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Letters to Biomet/EBI reps were sent on March 29, 2009 by US mail to alert them of the product recall and the affected parts that may be in their possession. They were also asked to notify each customer and locate all affected inventory. All letters advised of the recall and the specific lot numbers involved. A letter was sent to the hospital and physician that received affected parts and asked that implanting physicians make any medical decisions.
Recall details
- Recall number
- Z-2115-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 27, 2009
- Posted by the FDA
- Sep 21, 2009
- Recalling firm
- Ebi, L.P.
- Quantity
- 51
- Product code
- HSB
- FDA submissions
- K072161
- Distribution
- Worldwide Distribution -- USA and Spain.
Product description
Phoenix Cannulated Recon Screw; Ti-Cann PARTL THD Recon Screw Trauma Fixation Systems; Rx only Biomet Trauma, 100 Interpace Parkway, Parsippany, NJ 077054 Intended to aid in the alignment and stabilization of fractures to the skeletal system until healing has occurred.
Code information (models, lots, serial numbers)
Catalog number: 14-406110; Lot number 517350
Other devices under the same FDA submission
4 devices cleared under K072161 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PHOENIX NAIL SYSTEM 14-444181Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- PHOENIX RETROGRADE NAIL SYSTEM 14-444230Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- Phoenix Retrograde Nail System 14-444334Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
- Phoenix Retrograde Nail System 14-444342Biomet Orthopedics, LLC · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page