Skip to content

Arrow Multi-Lumen CVC w/ BlueFlex Tip (Central Venous… recallZ-1495-2009 · Arrow International Inc · initiated Feb 11, 2009

Terminated All recalls posted in June 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arrow International Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Leakage: Material not lipid resistant, may crack and/or leak. Use of this device with lipid containing solutions or 70% isopropyl alcohol may cause leakage or air embolism.

FDA root cause category: Device Design

What the company is doing

Arrow International issued a letter dated 2/18/09 to all customers informing them of the problem and if a lipid resistant stopcock is required to contact the firm for replacements. New labels state: Warning - do not use with lipid containing solutions or 70% isopropyl alcohol, as these solutions may cause leakage or air embolism.

Recall details

Recall number
Z-1495-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Feb 11, 2009
Posted by the FDA
Jun 24, 2009
Recalling firm
Arrow International Inc
Quantity
900,000 total, all products
Product code
DQY
FDA submissions
K862056
Distribution
Worldwide distribution: USA, Argentina, Australia, Bahamas, Belarus, Belgium, Brazil, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, Guatemala, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Singapore, Slovakia, South Africa, South Korea, Spain, Sri lanka, Sweden, Taiwan, Thailand, Tunisia, Turkey, UK, Ukraine, and Venezuela.

Product description

Arrow Multi-Lumen CVC w/ BlueFlex Tip (Central Venous Catheter)

Code information (models, lots, serial numbers)

All lots of Product codes listed in Z-1489-2009.

Other devices under the same FDA submission

132 devices cleared under K862056 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 51887).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page