Medtronic, Transfer Spike, Blood Processing Accessory, REF… recallZ-1424-2009 · Medtronic Cardiovascular Revascularization & Surgical Therap · initiated Feb 25, 2009
Terminated All recalls posted in June 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Cardiovascular Revascularization & Surgical Therap or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterility may be compromised. Non-vented caps that connect to the 3-way stopcock attached to the transfer spike could prevent Ethylene Oxide from contacting the inside surfaces of the stopcock during sterilization, leading to a potentially non-sterile product. Use of the product may cause serious health consequences from potential sepsis/infection and the associated remote risk of death.
FDA root cause category: Device Design
What the company is doing
Consignees were sent a Medtronic "Urgent Medical Device Recall Notice" letter dated March 5, 2009. The letter was addressed to Risk Manager. The letter described the product and problem and requested the consignee to complete, sign and return the certificate acknowledgement.
Recall details
- Recall number
- Z-1424-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Feb 25, 2009
- Posted by the FDA
- Jun 23, 2009
- Recalling firm
- Medtronic Cardiovascular Revascularization & Surgical Therap
- Quantity
- 20 units
- Product code
- DWE
- FDA submissions
- K800178
- Distribution
- Distribution to one medical facility in MN.
Product description
Medtronic, Transfer Spike, Blood Processing Accessory, REF BT945, Sequestration Kits
Code information (models, lots, serial numbers)
Lot 10279571, Use By 2012-01-31
Other devices under the same FDA submission
113 devices cleared under K800178 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA BB9L72R1Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS3 recalls0 reports
- NA BB8M16R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls1 report
- NA BB7P48R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7P46R4Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8L58R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8B70R7Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8D40R3Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB8T77R2Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA BB7V77RMedtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA BB7V76RMedtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA HY7S51R10Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
- NA BB8D56R9Medtronic, Inc. · TUBING, PUMP, CARDIOPULMONARY BYPASS1 recall1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page