Zimmer NexGen Complete Knee Solution Legacy Knee- Posterior… recallZ-1521-2009 · Zimmer Inc. · initiated Apr 17, 2009
Terminated All recalls posted in June 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The implant surface may not have been polished adequately, which could result in increased wear and polyethylene debris generation.
FDA root cause category: Employee error
What the company is doing
Consignees were notified by phone beginning on 4/17/09 and email and letter notifications were sent on 4/22/09. A dear surgeon letter entitled Urgent: Medical Device Recall, dated April 28, 2009, is being sent to implanting surgeons recommending patient monitoring. Consignees are to return the response form indicating their receipt of the notification. Questions or concerns should be directed to (800) 846-4637 or www.zimmer.com.
Recall details
- Recall number
- Z-1521-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 17, 2009
- Posted by the FDA
- Jun 25, 2009
- Recalling firm
- Zimmer Inc.
- Quantity
- 1
- Product code
- NJL
- FDA submissions
- P060037
- Distribution
- Worldwide Distribution -- USA including states of Arizona, California, Colorado, Hawaii, Illinois, Indiana, Michigan, Minnesota, Mississippi, New Jersey, New Mexico, New York, Ohio, Oregon, and Pennsylvania and country of Canada.
Product description
Zimmer NexGen Complete Knee Solution Legacy Knee- Posterior Stabilized LPS-FLEX Femoral Component Option Size D Left, sterile, Zimmer, Warsaw, IN; REF 00-5964-014-51.
Code information (models, lots, serial numbers)
Lot 61219714.
Other devices under the same FDA submission
26 devices cleared under P060037 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NexGen® Legacy® 5964-17-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing1 recall3 reports
- NexGen® Legacy® 5964-15-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-16-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-16-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing7 reports
- NexGen® Legacy® 5964-17-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing6 reports
- NexGen® Legacy® 5964-15-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing4 reports
- NexGen® Legacy® 5996-16-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® 5996-17-51Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® 5996-17-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing3 reports
- NexGen® Legacy® Prolong® 00-5917-050-10Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
- NexGen® Legacy® 5960-15-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
- NexGen® Legacy® 5964-13-52Zimmer, Inc. · Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Metal/Polymer, Mobile Bearing2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 52003).
- Z-1520-2009 · TerminatedZimmer NexGen Complete Knee Solution Legacy Knee - Posterior Stabilized LPS-Flex Femoral Component Precoat, Size D left , sterile, Zimmer, Warsaw, IN; REF 00-5960-014-51.
- Z-1522-2009 · TerminatedZimmer NexGen Complete Knee Solution Legacy Knee- Posterior Stabilized LPS-FLEX Femoral Component Option Size G Left , sterile, Zimmer, Warsaw, IN; REF 00-5964-017-51.
- Z-1523-2009 · TerminatedZimmer NexGen Complete Knee Solution Legacy Knee- Posterior Stabilized LPS Femoral Component Option Size D Left, sterile, Zimmer, Warsaw, IN; REF 00-5996-014-51.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page