Skip to content

Ortho Development Corp., Pivot BiPolar Head, 28mm, Hip… recallZ-0086-2010 · Ortho Development Corporation · initiated Apr 29, 2009

Terminated All recalls posted in October 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A Pivot BiPolar Head, the femoral head in a hip prosthesis, might not move freely after assembly due a stacking up of manufacturing tolerances.

FDA root cause category: Device Design

What the company is doing

Ortho sent a Field notification Memo dated 4/2/2009 to agents and distributors, giving them pre-implant examination guidelines. Decision to recall non-implanted units was made on 4/29/2009 and this was communicated by telephone to distributors.

Recall details

Recall number
Z-0086-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 29, 2009
Posted by the FDA
Oct 28, 2009
Recalling firm
Ortho Development Corporation
Quantity
23
Product code
KWY
FDA submissions
K050966
Distribution
AZ, CA, FL, MA, MI. No military, government or foreign distribution.

Product description

Ortho Development Corp., Pivot BiPolar Head, 28mm, Hip Prosthesis, Femoral Head.

Code information (models, lots, serial numbers)

Part 133-2846, Lots 3370, 3534, 3620, 3731 Part 133-2847, Lots 3492, 3535 Part 133-2848, Lot 3509 Part 133-2851, Lot 3433 Part 133-2852, Lot 3661

Other devices under the same FDA submission

10 devices cleared under K050966 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page