Skip to content

Biomet Biolox-Delta Modular Ceramic Head, 28 mm head… recallZ-1613-2009 · Biomet, Inc. · initiated Mar 16, 2009

Terminated All recalls posted in July 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has received reports of fractured heads, requiring revision surgery.

FDA root cause category: Process control

What the company is doing

Biomet Orthopedics issued an "Urgent Medical Device Recall Notice" dated March 16, 2009 to users describing the affected device. Consignees were instructed to locate and discontinue use of the product and quarantine until a sales representative can arrange for removal. Consignees were also asked to complete and return the attached "Fax-Back Response Form" to Biomet at 1-574-372-1683. Further questions can be directed to Biomet Orthopedics at 1-800-348-9500.

Recall details

Recall number
Z-1613-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 16, 2009
Posted by the FDA
Jul 14, 2009
Recalling firm
Biomet, Inc.
Quantity
174
Product code
LZO
FDA submissions
K042091
Distribution
Worldwide Distribution -- US including PR and states of AZ, AL, AR, CA, CO, FL, IL, IN, KY, KS, MA, MI, LA, NY, SD, GA, MN, PA, NJ, MD, MI, OH, ME, TX, WA, WI and VA and countries of Australia, Belgium, Canada, Costa Rica, Finland, Netherlands and South Wales.

Product description

Biomet Biolox-Delta Modular Ceramic Head, 28 mm head diameter, plus 5 neck, type 1, taper, sterile, Biomet Orthopedics, Inc., Warsaw, IN; Catalog Number 12-115112. A hip joint metal/ceramic/polymer semiconstrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint.

Code information (models, lots, serial numbers)

Lots 114380, 175490, 229000, 310710, 342670, 385840, 403940, 467870, 484970, 572770, 641430, 705380, 761420 and 792280.

Other devices under the same FDA submission

8 devices cleared under K042091 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 52092).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page