Skip to content

Boston Scientific, Fort Guide Wire, Floppy, 300cm x 2 cm… recallZ-1487-2009 · Boston Scientific Corporation · initiated May 12, 2009

Terminated All recalls posted in June 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product packaged incorrectly. One batch/lot of wires, that should contain radiopaque marker bands, is missing the marker bands; and a second batch/lot of wires, which should not contain radiopaque marker bands, has the marker bands. Since the marker bands issue would not be visually apparent to the physician there is a potential for prolongation or delay of the procedure in order to exchange the

FDA root cause category: Process control

What the company is doing

Consignees were sent a "Boston Scientific Urgent Medical Device Recall" letter dated may 19, 2009. The letter was addressed to "Dear Risk Manger/ Field Action Contact". The letter described the problem, customer steps for recall and listing of product batches / lots. The letter requested consignees to return the Reply Verification Tracking Form.

Recall details

Recall number
Z-1487-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 12, 2009
Posted by the FDA
Jun 24, 2009
Recalling firm
Boston Scientific Corporation
Quantity
Total of 4 boxes (5 units /box).
Product code
DQX
FDA submissions
K011968
Distribution
USA distribution: AR, MO, CA, FL, and NV.

Product description

Boston Scientific, Fort Guide Wire, Floppy, 300cm x 2 cm REF 34949-04, for use in intravascular interventional procedures.

Code information (models, lots, serial numbers)

01403876

Other devices under the same FDA submission

2 devices cleared under K011968 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 52098).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page