Alaris Pump Module, Model 8100 recallZ-1793-2009 · Cardinal Health 303 dba Cardinal Health · initiated Jun 12, 2009
Terminated All recalls posted in August 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 303 dba Cardinal Health or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential risk of over infusion when certain fluids solidify in the pumping mechanism of the Alaris Pump module. This notification is to reinforce the importance of proper techniques for IV set loading and cleaning. When IV sets are improperly loaded, the set can tear and cause medication to leak and potentially enter the bezel area. Improper cleaning of the device, such as using hig
FDA root cause category: Device Design
What the company is doing
The recall was initiated on June 12, 2009 with the firm forwarding via UPS an Urgent: Medical Device Recall Notification with attached Customer Response Card to the Chief Administrative Officer, Director of Pharmacy, Director of Biomedical Engineering, Director of Nursing & Director of Risk Management of all of the affected consignees. Customers were also asked to please promptly complete and return the enclosed Customer Response Card to expedite the correction process. The Recall Notification informed the consignees of: (A) PROBLEM #1: Occlusion Warning Message for Alaris Pump Module; (B) PROBLEM #2 Syringe volume Warning Message for Alaris PCA module; (C) PROBLEM #3 Electrostatic Discharge (ESD) protection circuitry; (D) PROBLEM #4 Fluid Ingress and (E) PROBLEM #5 Alaris System Inter-Unit Interface (IUI) Connectors Labeling. The recall notice included "Required Action for Users & Follow-up actions by Cardinal Health for all five problems. Customers were also asked to report any adverse reactions experienced with the use of this product, and/or quality problems to the FDA's MedWatch Program. In addition, customer were provided the following contact information for the recall: (1) Cardinal Health Recall Center for general questions: By phone at 1-888-562-6018 7am to 5pm (Pacific) ; (2) Customer Advocacy for adverse reports: By phone at 1-800-854-7120, Option 1, Option 1, Option 3 or by email at [email protected] 24 hours a day, Sunday-Saturday; (3) Technical Support for Technical Questions regarding the Alaris System: By phone at 1-888-562-6018, 6am-5pm (Pacific). The firm followed-up with the release of a nationwide press statement regarding this issue on July 29, 2009. The press release indicated serial numbers of affected devices, as well as CareFusions short term instructions to customers, and the firms strategy to fix the affected devices can be found at: www.cardinalhealth.com/alaris/medical-device-recall/ (This informatio
Recall details
- Recall number
- Z-1793-2009
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jun 12, 2009
- Posted by the FDA
- Aug 3, 2009
- Recalling firm
- Cardinal Health 303 dba Cardinal Health
- Quantity
- 300,000 units
- Product code
- FRN
- FDA submissions
- K950419
- Distribution
- Worldwide Distribution -- US (including Puerto Rico and Guam), Germany, England, Canada, Netherlands, Malaysia, Singapore, Australia, and New Zealand.
Product description
Alaris Pump Module, Model 8100.
Code information (models, lots, serial numbers)
Model 8100.
Other devices under the same FDA submission
1 device cleared under K950419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Alaris 8100Carefusion 303, Inc. · Pump, infusion2 recalls2 deaths650,218 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 52119).
- Z-1790-2009 · TerminatedThe Alaris PC Unit, Model Numbers 8000 and 8015 (formerly Medley PC Unit) with software version 4 and above.
- Z-1791-2009 · TerminatedThe Alaris PC Unit, Model Numbers 8000 and 8015 (formerly Medley PC Unit) software version 8 and above when used with the Alaris PCA Module (Model 8120). The Alaris PC Unit is the…
- Z-1792-2009 · TerminatedThe Alaris PC Unit, Model 8015 (manufactured or serviced between April 2008 and January 2009). The Alaris PC Unit is the central programming and power supply component for the Ala…
- Z-1794-2009 · TerminatedAlaris System Inter-Unit Interface (IUI) Connectors. The Modules include the following: Alaris PC Unit, Model Numbers 8000 and 8015; Alaris Pump Module, Model 8100; Alaris Patien…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page