Tibial Tray, component of Ortho Development Balanced Knee… recallZ-1993-2009 · Ortho Development Corporation · initiated May 29, 2009

Terminated All recalls posted in September 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho Development Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Erroneous lot number on 3 of 28 units of Balanced Knee System. Three units from Lot Number 0055245 were mixed in with the recalled lot (0054891).

FDA root cause category: Mixed-up of materials/components

What the company is doing

Consignees were notified via e-mail on May 29, 2009 and June 5, 2009 requesting return of units. Direct questions about the recall to the Ortho Development Corporation by calling 1-801-619-3450.

Recall details

Recall number
Z-1993-2009
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 29, 2009
Posted by the FDA
Sep 9, 2009
Recalling firm
Ortho Development Corporation
Quantity
13 units
Product code
JWH
FDA submissions
K994370
Distribution
Worldwide -- US (states of FL and TX) and Japan.

Product description

Tibial Tray, component of Ortho Development Balanced Knee System, Nonporous, Sterile R, Size 3, REF 162-1300, Ortho Development, Draper, Utah 84020. The Balanced Knee System is indicated in the salvage of previously failed surgical attempts where femoral bone loss does not require the use of augments or stem extensions and where collateral ligaments may be relied upon for medial/lateral stability.

Code information (models, lots, serial numbers)

Lot Number 0054891.

Other devices under the same FDA submission

4 devices cleared under K994370 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates