Oxinium Femoral Head, 36 MM OD-3, 12/14 Taper, REF… recallZ-0389-2010 · Smith & Nephew Inc · initiated Jan 23, 2009
Terminated All recalls posted in November 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Radius at the end of the taper of the femoral implant is smaller than specification. This error could result in the tapers not locking fully or properly on femoral stems or in causing impingement and/or increased wear debris. The deficiency may not be noticed by the user. Impingement and/or wear debris may lead to revision.
FDA root cause category: Other
What the company is doing
All affected Smith & Nephew Sales Representatives were notified of problem and the recall overnight mail on 01/23/2009. All affected Smith & Nephew International Distributors were notified via email and telephone on 01/23/2009. They were instructed to immediately quarantine the product for return. Further questions are directed to the contacts at Smith and Nephew listed in the letter.
Recall details
- Recall number
- Z-0389-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Jan 23, 2009
- Posted by the FDA
- Nov 20, 2009
- Recalling firm
- Smith & Nephew Inc
- Quantity
- 925 units
- Product code
- JDI
- FDA submissions
- K030766
- Distribution
- Nationwide; 17 international: Switzerland, Dubai, Brazil, Canada, South Africa, Australia, Belgium, Spain, Thailand Turkey, Costa Rica, Denmark, India, Italy, and France.
Product description
Oxinium Femoral Head, 36 MM OD-3, 12/14 Taper, REF 71343603, Sterile, Smith & Nephew, Memphis TN 38116
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Lot Numbers: 08BM09885, 08EM06396, 08FM14910, 08HM10480, 08KM07754, 08KM12526, 08HM03206, 08GM03007R, 08EM19444, 08EM19445, 08EM19446, 08EM20186, 08FM00231, 08RM00242, 08FM03526, 08FM04628, 08FM05689, 08FM06247, 08FM14906, 08FM17347, 08FM22085, 08FM22087, 08GM03007, 08GM03489, 08GM06660, 08GM11382, 08GM11425, 08GM12094, 08GM13061, 08GM17805, 08GM18609, 08HM02631, 08HM03926, 08HM10037, 08HM13107, 08HM13118, 08HM16147, 08HM16150, 08HM17621, 08HM18516, 08JM00370, 08JM04970, 08JM04971, 08JM05505, 08JM05506, 08JM08122, 08JM09041, 08JM11262, 08JM14571, 08JM15417, 08JM19246, 08JM19247, 08JM19248, 08JM20328, 08JM21716, 08JM21717, 08JM21718, 08KM03713, 08KM07732, 08KM14807, 08KM20680, 08KM23716, 08LM01328, 08LM02300, 08LM11759, 08LM12842, and 08LM13924
Other devices under the same FDA submission
12 devices cleared under K030766 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- OXINIUM 71343600Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented81 reports
- OXINIUM 71343200Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented58 reports
- OXINIUM 71343604Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented53 reports
- OXINIUM 71343603Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented41 reports
- OXINIUM 71343204Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented38 reports
- OXINIUM 71342800Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented35 reports
- OXINIUM 71343608Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented33 reports
- OXINIUM 71342804Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented29 reports
- OXINIUM 71343203Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented28 reports
- OXINIUM 71343208Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented22 reports
- OXINIUM 71342808Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented15 reports
- OXINIUM 71342200Smith & Nephew, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented13 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page