Vigilance II Monitor Model: VIG2/VIG2E Software: versions… recallZ-0942-2012 · Edwards Lifesciences, Llc · initiated Sep 12, 2008

Terminated All recalls posted in February 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Edwards Lifesciences, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Edwards Lifesciences has made the decision to initiate a field corrective action for the Edwards Vigilance II Model: Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73 due to the potential of inaccurate Injectate Cardiac Output (ICO) measurements.

FDA root cause category: Device Design

What the company is doing

Edwards initiated two Urgent Product Notification - Action Required letters to all affected customers. The letters identified the affected product, problem, and actions to be taken. The letters instructed users to immediately discontinue use of in-line temperature probes for Intermittent Cardiac Output (ICO) measurements on the Vigilance II instrument and recommends use of the bath probe only (model 9850A) to measure the temperature of the injectate solution for ICO calculations, until software update have been completed. Customers were instructed to complete and return a software upgrade form to Edwards as record of completion. For questions call Edwards Technical Support department at 800-822-9837.

Recall details

Recall number
Z-0942-2012
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Sep 12, 2008
Posted by the FDA
Feb 2, 2012
Recalling firm
Edwards Lifesciences, Llc
Quantity
6934 instruments
Product code
DXG
FDA submissions
K043103
Distribution
Worldwide Distribution - Nationwide Distribution (USA) - and the countries of: Argentina, Australia/ New Zealand, Brazil, Canada, China, Columbia, Ecuador, Europe, Indonesia, Japan, Korea, Macao, Malaysia, Mexico, Middle East, Peru, Puerto Rico, Singapore, Thailand, Taiwan and Venezuela

Product description

Vigilance II Monitor Model: VIG2/VIG2E Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73. Product Usage: The Vigilance II CCO/SvO2/CEDV Monitor is indicated for use in patients requiring monitoring of hemodynamic parameters, including cardiac output, oximetry and right ventricular, ejection, fraction, and end diastolic volume.

Code information (models, lots, serial numbers)

Software: versions 00.51, 00.60, 00.61, 00.62, 00.63, 00.64, 00.67, 00.69, 00.71, 00.73.  Serial numbers: VG000031 to VG007468

Other devices under the same FDA submission

1 device cleared under K043103 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.