Zimmer Gender Solutions Natural-Knee Flex System, Prolong… recallZ-0239-2010 · Zimmer Inc. · initiated Oct 8, 2009
Terminated All recalls posted in November 2009
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The surgeon may have difficulty inserting the polyethylene into the tibial baseplate, which may result in intraoperative damage to the device.
FDA root cause category: Other
What the company is doing
An "Urgent: Field Correction" dated October 8, 2009 was issued to Consignees. The notification letter described the affected product, issue, health risks and actions for customers to take. The letter notified consignees of a changed to the surgical technique, to include the use of a new peripheral rib impactor tool. Notifications of this field notification were sent to all affected direct accounts of Zimmer, Inc. Contact your Zimmer Sales Representative if you have any questions regarding this action. For shipping assistance, questions or assistance in notifying your accounts about the field correction please contact Zimmer, Inc. at 1-800-613-6131.
Recall details
- Recall number
- Z-0239-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 8, 2009
- Posted by the FDA
- Nov 13, 2009
- Recalling firm
- Zimmer Inc.
- Quantity
- 26,404 of all products
- Product code
- JWH
- FDA submissions
- K070214
- Distribution
- Worldwide Distribution -- United States, Austria, Belgium, France, Germany, Japan, Spain and Switzerland.
Product description
Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, ultracongruent, Size 3,4,5, left, 9 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Warsaw, IN; REF 00-5428-013-09. The device is a semiconstrained total knee prosthesis components indicated for uncemented or cemented use in skeletally mature individuals.
Code information (models, lots, serial numbers)
Lots 60932788, 60937764, 60937765, 60958306, 60981870, 60988244, 60993007, 60993008, 60993009, 61001563, 61005458, 61017819, 61017836, 61036379, 61036409, 61045107, 61045108, 61051489, 61065060, 61069428, 61069443, 61075263, 61078772, 61078797, 61096381, 61096382, 61101484, 61106577, 61114197, 61114198, 61117066, 61122290, 61126556 and 61136945.
Other devices under the same FDA submission
14 devices cleared under K070214 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Gender Solutions® Natural-Knee® Prolong® 00-5420-008-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER3 reports
- Gender Solutions® Natural-Knee® 00-5414-016-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Gender Solutions® Natural-Knee® Prolong® 00-5420-008-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- Gender Solutions® Natural-Knee® Prolong® 00-5420-008-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Gender Solutions® Natural-Knee® Prolong® 00-5428-021-09Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 reports
- Gender Solutions® Natural-Knee® Prolong® 00-5428-023-09Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- Gender Solutions® Natural-Knee® 00-5410-017-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Gender Solutions® Natural-Knee® 00-5414-014-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Gender Solutions® Natural-Knee® Prolong® 00-5420-010-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Gender Solutions® Natural-Knee® Prolong® 00-5428-011-09Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- Gender Solutions® Natural-Knee® Prolong® 00-5428-011-11Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER1 report
- Gender Solutions® Natural-Knee® Prolong® 00-5428-011-13Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 53399).
- Z-0199-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, left, 9 MM height, use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN; …
- Z-0200-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, left, 11 MM height, use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN;…
- Z-0201-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, left, 13 MM height, use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN;…
- Z-0202-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, left, 16 MM height, use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN;…
- Z-0203-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, left, 19 MM height, use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN;…
- Z-0204-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 9 MM height, use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, IN; RE…
- Z-0205-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 11 MM height, use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, IN; R…
- Z-0206-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 13 MM height, for use with baseplate size 1 or 2 sterile, Zimmer, Warsaw, I…
- Z-0207-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 16 MM height, for use with baseplate size 1 or 2 sterile, Zimmer, Warsaw, I…
- Z-0208-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, left, 19 MM height, for use with baseplate size 1 or 2 sterile, Zimmer, Warsaw, I…
- Z-0209-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, left, 9 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Warsa…
- Z-0210-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, left, 11 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Wars…
- Z-0211-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, left, 13 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Wars…
- Z-0212-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, left, 16 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Wars…
- Z-0213-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, left,19 MM height, for use with baseplate size 3, 4 or 5, sterile, Zimmer, Warsa…
- Z-0214-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 9 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw,…
- Z-0215-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 11 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw…
- Z-0216-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 13 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw…
- Z-0217-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 16 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw…
- Z-0218-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 19 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw…
- Z-0219-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1 or 2, right, 9 MM height, for use with baseplate size 1, 2, sterile, Zimmer, Warsaw, …
- Z-0220-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, right, 11 MM height, for use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, …
- Z-0221-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, right, 13 MM height, for use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, …
- Z-0222-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, right, 16 MM height, for use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, …
- Z-0223-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 1,2, right, 19 MM height, for use with baseplate size 1 or 2, sterile, Zimmer, Warsaw, …
- Z-0224-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, right, 9 MM height, for use with baseplate size 3, 4, or 5, sterile, Zimmer, War…
- Z-0225-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, right, 11 MM height, for use with baseplate size 3, 4, or 5, sterile, Zimmer, Wa…
- Z-0226-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, right, 13 MM height, for use with baseplate size 3, 4, or 5, sterile, Zimmer, Wa…
- Z-0227-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, right, 16 MM height, for use with baseplate size 3, 4, or 5, sterile, Zimmer, Wa…
- Z-0228-2010 · TerminatedZimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 3,4,5, right, 19 MM height, for use with baseplate size 3, 4, or 5, sterile, Zimmer, Wa…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page