Skip to content

Richard Wolf Hulka Clip recallZ-0272-2010 · Richard Wolf Medical Instruments Corp. · initiated Oct 16, 2009

Terminated All recalls posted in November 2009

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Richard Wolf Medical Instruments Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The sterile pouch seal may fail to remain intact during handling rendering the Hulka clip non-sterile.

FDA root cause category: Packaging

What the company is doing

Richard Wolfe sent recall letters dated October 16, 2009 to the direct accounts via certified mail on the same date. The accounts were informed that the lot of Hulka clips did not meet their sterile seal burst test requirements, and that the sterile seal may fail to remain intact during handling, resulting in the clips becoming contaminated or non-sterile. They were instructed to inspect their inventory for the affected lot, quarantine any product found and call Richard Wolf at 800-323-9653, ext 225 or 380 to arrange for the return of the affected product for replacement. The accounts were also requested to complete and return the enclosed response form to Richard Wolf, even if they do not have any of the affected lot in stock.

Recall details

Recall number
Z-0272-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 16, 2009
Posted by the FDA
Nov 17, 2009
Recalling firm
Richard Wolf Medical Instruments Corp.
Quantity
150 cartons
Product code
HGB
FDA submissions
P870080
Distribution
Nationwide

Product description

Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure tray/pouch, 10 pouches per shipping carton; Richard Wolf Medical Instruments Corp., Vernon Hills, IL 60061; item 4986.90. Indicated for use for female sterilization (permanent contraception) by occluding the fallopian tubes

Code information (models, lots, serial numbers)

lot number 1391, expiry date 02.2010

Other devices under the same FDA submission

1 device cleared under P870080 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page