Zimmer NexGen Complete Knee Solution Minimally Invasive… recallZ-1064-2010 · Zimmer Inc. · initiated Nov 16, 2009
Terminated All recalls posted in March 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The titanium portion of the implant may separate from the trabecular metal material.
FDA root cause category: Other
What the company is doing
On December 1, 2009, an e-mail was sent to consignees for lot 60866613 to quarantine and return the product. On December 2, 2009 an e-mail was sent to U.S. distributors, and on December 3, 2009 and e-mail was sent to international distributors, instructing them to suspend usage and return all lots of size 8 trays. On December 9, 2009, a recall letter was sent to U.S. user accounts who received the products directly from Zimmer, U.S. distributors and Zimmer's U.S. sales force informing them of the recall. Direct questions about this recall to Zimmer, Inc. by calling 1-800-613-6131.
Recall details
- Recall number
- Z-1064-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Nov 16, 2009
- Posted by the FDA
- Mar 10, 2010
- Recalling firm
- Zimmer Inc.
- Quantity
- 846
- Product code
- JWH
- FDA submissions
- K072160
- Distribution
- Worldwide Distribution -- United States, Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Saudi Arabia, Spain, Sweden, Switzerland and the United Kingdom.
Product description
Zimmer NexGen Complete Knee Solution Minimally Invasive Solutions Procedures Trabecular Metal Technology Tibial Tray, fixed bearing, size 8, sterile, Zimmer, Warsaw, Indiana; REF 00-5954-057-02. The product is used as an implant for knee replacement.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
All lots. Lots 00111134, 60747593, 60747594, 60752676, 60752677, 60760248, 60768076, 60821507, 60824004, 60830411, 60830412, 60845351, 60850604, 60852670, 60852671, 60860895, 60866613, 60872311, 60879050, 60892605, 60900745, 60907449, 60917234, 60922494, 60938889, 60938890, 60947027, 60947028, 60963035, 60963036, 60963037, 60963039, 60983254, 61015851, 61028168, 61033723, 61033724, 61068387, 61068388, 61068389, 61074754, 61074755, 61084771, 61096604, 61116246, 61116247, 61137056, 61143427, 61149041, 61157940, 61173295, 61193516, 61193520, 61198179, 61204038, 61215644, 61225562, 61235612, 61240323, 61248321, 61259184, 61263170, 61272449, 61285508, 61298201, 61306497, 61280371, 61315873, 61319581 and 61329266.
Other devices under the same FDA submission
5 devices cleared under K072160 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NexGen® Trabecular Metal® 00-5954-047-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER3 reports
- NexGen® Trabecular Metal® 00-5954-037-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- NexGen® Trabecular Metal® 00-5954-037-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, COATED, POLYMER/METAL/POLYMER2 reports
- NexGen® Trabecular Metal® 00-5954-047-02Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
- NexGen® Trabecular Metal® 00-5954-057-01Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page