Skip to content

Powerlink System with IntuiTrak Delivery System… recallZ-0593-2010 · Endologix Inc · initiated Nov 25, 2009

Terminated All recalls posted in January 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Endologix Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

This action is being taken because the product has the potential for the polyimide tubing to detach from the inner core due to the bond failure. If this occurs, treatment may be delayed or interrupted. The firm's initial health hazard analysis indicated the event may cause temporary or medically reversible adverse health consequences or an outcome where the probability of serious adverse health c

FDA root cause category: Process design

What the company is doing

The recall communication was initiated on 11/25/09 by phone call and followed by recall letter via Fed-Ex Overnight. The recall letter informed the consignees of the reason for recall and products affected. Users were instructed to share this information with physicians that performed the affected procedures at their facility. Endologix informed their consignees that their personnel would retrieve the affected devices and return it to Endologix at no cost and would replace the affected units. Consignees with questions or comments about the notice are directed to (800) 983-2284.

Recall details

Recall number
Z-0593-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 25, 2009
Posted by the FDA
Jan 11, 2010
Recalling firm
Endologix Inc
Quantity
3 UNITS DISTRIBUTED: 1 LOT W09-2966, 2 LOT W09-2998
Product code
OAD
FDA submissions
P040042
Distribution
Recalled devices were distributed to medical centers and hospitals in OR, PA, FL, TX, MI and LA.

Product description

Powerlink System with IntuiTrak Delivery System (Endovascular Graft): Model Number 34-34-80L, LOTS W09-2966 AND W09-2998.

Code information (models, lots, serial numbers)

MODEL NUMBER 34-34-80L, LOTS W09-2966 AND W09-2998.

Other devices under the same FDA submission

3 devices cleared under P040042 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 53983).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page