Skip to content

Elekta Synergy XVI Medical Linear Accelerator. Model… recallZ-1405-2010 · Elekta, Inc. · initiated Dec 12, 2008

Terminated All recalls posted in April 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

It is possible that two (or more) CT reference data sets are necessary for one patient, for example, if you have to scan a patient during breath hold and then again during free breathing.

FDA root cause category: Software design

What the company is doing

Elekta, Inc. issued an "Important Notice" dated December 12, 2008. Customers were informed of the affected device and instructed on proper steps until an update is provided. For further information, contact Elekta, Inc. at 1-770-300-9725.

Recall details

Recall number
Z-1405-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 12, 2008
Posted by the FDA
Apr 16, 2010
Recalling firm
Elekta, Inc.
Quantity
76 units
Product code
IYE
FDA submissions
K051932
Distribution
Nationwide Distribution -- AL, AZ, CO, CT, FL, GA, ID, KS, KY, MA, MI, MN, MS, NE, NY, OH, OK, OR, PA, TN, TX, VA, WA and WV.

Product description

Elekta Synergy XVI Medical Linear Accelerator. Model Numbers R3.5, R4.0 and R4.2. Elekta, Inc. Norcross, GA 30092. Intended for use as a radiation therapy treatment of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Code information (models, lots, serial numbers)

Serial Numbers: 151008, 151024, 151025, 151033, 151038, 151051, 151053, 151055, 151059, 151073, 151074, 151076, 151080, 151084, 151093, 151104, 151110, 151130, 151134, 151141, 151143, 151151, 151154, 151155, 151156, 151157, 151160, 151167, 151168, 151173, 151176, 151178, 151216, 151228, 151229, 151238, 151245, 151250, 151253, 151256, 151259, 151260, 151266, 151267, 151269, 151298, 151301, 151305, 151309, 151317, 151325, 151326, 151327, 151328, 151340, 151352, 151359, 151360, 151362, 151363, 151369, 151371, 151387, 151394, 151398, 151411, 151416, 151421, 151422, 151423, 151430, 151433, 151435, 151441, 151446 and 151479.

Other devices under the same FDA submission

1 device cleared under K051932 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page