Alaris Pump Module model 8100 recallZ-2185-2012 · CareFusion 303, Inc. · initiated Jun 29, 2012

Class I Terminated All recalls posted in August 2012

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The recall was initiated because Carefusion identified a potential risk associated with the Alaris Pump module model 8100. Carefusion has received reports of customers experiencing motor stalls during infusion with Alaris Pump Module (model 8100) manufactured between August 2010 and July 2011.

FDA root cause category: Under Investigation by firm

What the company is doing

CareFusion sent an Urgent Medical Device Recall Notification letter dated July 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Required action for users instructed high risk patients undergoing infusions at high rates, consider having additional devices as back up. Clinicians should weigh the risk/benefit to the patients before continuing to use the device. Customers with recall related questions were instructed to contact Carefusion support center at (888) 562-6018. Customers with adverse event report questions were instructed to contact customer advocacy at (800) 854-7128, option 1, option 1, option 3 or email at [email protected]. Customers with technical questions regarding the Alaris system were instructed to contact Technical support at (888) 812-3229. For questions regarding this recall call 858-617-4000.

Recall details

Recall number
Z-2185-2012
Recall class
Class I (most serious)
Status
Terminated
Initiated by the company
Jun 29, 2012
Posted by the FDA
Aug 15, 2012
Recalling firm
CareFusion 303, Inc.
Quantity
102,248 units
Product code
FRN
FDA submissions
K950419
Distribution
Worldwide Distribution - USA (nationwide) and the countries of: South Africa and Canada. Additional Countries: Netherlands.

Product description

Alaris Pump Module model 8100. Subsequent product code: FPA The Pump module is intended for facilities that utilize infusion for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery through clinically acceptable routes of administration; such as, intravenous (IV), intra-arterial (IA),subcutaneous, epidural, enteral, or irrigation of fluid spaces. The Pump module is indicated for use on adults, pediatrics, and neonates.

Code information (models, lots, serial numbers)

All serial numbers associated with the Alaris Pump Module Model 8100: 8100 DEMO PUMP MODULE; 8100 Global Pump Module; 8100 IOC Pump Module V8.5.29.0; 8100 LVP Training Unit; 8100 Pump Module V8.5.29.0; 8100 Pump Module V8.5.29.0 (V9); Medley Patient Care System LVP; Model 8100 Trade-In Material; RCND 8100 Pump Module V8.5.29.0

Other devices under the same FDA submission

1 device cleared under K950419 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.