Skip to content

ARCHITECT HAVAB-M Reagent recallZ-1596-2010 · Abbott Diagnostics International, Ltd. · initiated Dec 30, 2009

Terminated All recalls posted in May 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Diagnostics International, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential to generate a false grayzone or reactive HAVAB-M result when the ARCHITECT AUSAB assay precedes the ARCHITECT HAVAB-M assay.

FDA root cause category: Pending

What the company is doing

Direct accounts (hospitals, private laboratories, and other institutions) were sent a Product Correction letter, dated 12/30/2009. The letter identified the affected product, explained the issue and patient impact, and listed several necessary actions. Customers were instructed to follow one of 4 different alternatives to protect the integrity of test results. The letter is to be retained for laboratory records and a copy should be sent to other laboratories that kits were forwarded to. Questions regarding this information should be directed to Customer Service at 1-877-4ABBOTT.

Recall details

Recall number
Z-1596-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 30, 2009
Posted by the FDA
May 19, 2010
Recalling firm
Abbott Diagnostics International, Ltd.
Quantity
1,326
Product code
LOL
FDA submissions
K063329
Distribution
Nationwide Distribution -- Including Puerto Rico.

Product description

ARCHITECT HAVAB-M Reagent A chemiluminescent microparticle immunoassay (CMIA) for the qualitative detection of IgM antibody to hepatitis A virus (IgM anti-HAV) in human adult and pediatric serum and plasma (dipotassium EDTA, lithium heparin, and sodium heparin) and neonatal serum.

Code information (models, lots, serial numbers)

Lot # 76175Q100 exp. date 4/18/10; Lot # 79412Q100 exp. date 7/22/10; Lot # 82584Q100 exp. date 10/20/10.

Other devices under the same FDA submission

1 device cleared under K063329 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page