Why it was recalled

The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.

FDA root cause category: Other

What the company is doing

Abbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.

Recall details

Recall number
Z-0456-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Mar 17, 2010
Posted by the FDA
Nov 24, 2010
Recalling firm
Abbott Point of Care Inc.
Quantity
34525 cartridges
Product code
CGL
FDA submissions
K973292
Distribution
Worldwide Distribution: Throughout USA and Canada.

Product description

i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.

Code information (models, lots, serial numbers)

Lot numbers A09279A and A09280A, Exp March 2010

Other devices under the same FDA submission

1 device cleared under K973292 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.