Skip to content

directCHECK Whole Blood Control recallZ-1786-2010 · International Technidyne Corporation · initiated Apr 15, 2010

Terminated All recalls posted in June 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from International Technidyne Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Indication not cleared or approved. The acceptable performance range published in the directCHECK Microcoagulation Quality Control package insert is incorrect.

FDA root cause category: Error in labeling

What the company is doing

Notification letters were sent vis Federal Express on April 15, 2010 to all distributors and end users to return unused product or to replace the package insert. Please direct questions ITC Technical Support at phone 800-631-5945 Ext. 4707.

Recall details

Recall number
Z-1786-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 15, 2010
Posted by the FDA
Jun 11, 2010
Recalling firm
International Technidyne Corporation
Quantity
94 boxes
Product code
GGN
FDA submissions
K944691
Distribution
Distributors and medical facilities in AZ, CA, IL, KY, MA, MD, ME, MO, PA, WA. Foreign distribution to Australia, Canada, Switzerland, Germany, Spain, Quatar, and Saudi Arabia.

Product description

directCHECK Whole Blood Control; for Hemochron J. Microcoagulation Systems.

Code information (models, lots, serial numbers)

Catalog number.Code number: DCJPT-A; Lot number G9DPA005; Exp. 1/31/2011.

Other devices under the same FDA submission

1 device cleared under K944691 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page