Skip to content

The Dimension TNI Flex reagent cartridge (Cardiac… recallZ-0460-2011 · Siemens Healthcare Diagnostics, Inc. · initiated Apr 12, 2010

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Some LOCI cartridges have sub-optimal seal of the lidstock that can cause rupture of the package at high elevations, causing the cartridge to leak. Use of leaking cartridges may result in inaccurate test results.

FDA root cause category: Other

What the company is doing

Siemens issued an "Urgent Field Safety Notice" dated April 2010, informing users of the problem and instructing them to inspect cartridges for leaking fluid. Cartridges with leaking fluid should be reported to the Siemens Technical Solutions Center. The Technical Solutions Center can be contacted at 800 441 9250.

Recall details

Recall number
Z-0460-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 12, 2010
Posted by the FDA
Nov 24, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
4966
Product code
MMI
FDA submissions
K081643
Distribution
Worldwide Distribution: USA, Europe, Australia, China, Japan, Korea, and Saudi Arabia.

Product description

The Dimension TNI Flex reagent cartridge (Cardiac Troponin), manufactured by Siemens Healthcare Diagnostics, Inc., Newark, DE.

Code information (models, lots, serial numbers)

Lot# FD0110 Exp# 04/20/2010, Lot#FD0141, EXP# 05/21/2010, Lot# ED0155, EXP# 06/04/2010, Lot# EA0227, EXP# 08/15/2010, Lot# EB0247, EXP#09/04/2010, Lot# FA2096, EXP#10/23/2010, Lot#FC0310, EXP# 11/06/2010, Lot# FA0330, EXP# 11/26/2010.

Other devices under the same FDA submission

2 devices cleared under K081643 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 55604).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page