Hospital Grade Power Cord, UL, Part # 66319, 055476… recallZ-1939-2010 · Medtronic Cardiovascular Revascularization & Surgical Therap · initiated Apr 21, 2010
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Cardiovascular Revascularization & Surgical Therap or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The potential risks during the use of this power cord include electrical shock, delay in setup and therapy, interruption of therapy, device failure, and fires. Although Medtronic has received no reports of adverse events related to the affected cord, a failure may potentially lead to serious adverse health consequences, depending on the device, and therapy being interrupted. This Field Notificatio
FDA root cause category: Nonconforming Material/Component
What the company is doing
Consignees were sent on 4/21/10 a Medtronic "Urgent Medical Device Recall Notice" letter dated April 19, 2010. The letter was addressed to their customers and described the problem and the product involved. They recommended to discontinue use of the AC power cords and to destroy them. Requested consignees to fill out the attached Medtronic Recall Certificate (FCA 1005), and return it to Medtronics.
Recall details
- Recall number
- Z-1939-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Apr 21, 2010
- Posted by the FDA
- Jun 30, 2010
- Recalling firm
- Medtronic Cardiovascular Revascularization & Surgical Therap
- Quantity
- 619 USA, 80 OUS
- Product code
- DWA
- FDA submissions
- K051303, K941921, K972894
- Distribution
- Worldwide distribution: USA, Latin America, Middle East, Japan, and Canada.
Product description
Hospital Grade Power Cord, UL, Part # 66319, 055476, manufactured by Electri-Cord Manufacturing Co. beginning in February 2003. The cords are a detachable unit, supplied with the instruments listed below: Medtronic Bio-Console 550 centrifugal pump console, Model #95180, Catalogue # 95180, E95180, R95180. Medtronic Bio-Console 560 centrifugal pump consoles Model # 560BC, Catalogue # 560BC, 560BCS, 560BCS1. Medtronic autoLog, Model # ATLG110, Catalogue # ATLG110, ATLG110E ATLG110R. Medtronic Sequestra 1000, Model #SEQ1000, Catalogue # SEQ1000, SEQ1000E, SEQ1000R.
Code information (models, lots, serial numbers)
Part # 66319 & 055476
Other devices under the same FDA submission
3 devices cleared under K051303, K941921, K972894 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA ATL2001Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION9 reports
- autoLog® ATLG110Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION2 reports
- autoLog® ATLG110RMedtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page