STERIS 4085, 5085 & 5085SRT Surgical Tables recallZ-1941-2010 · Steris Corporation · initiated May 14, 2010
Terminated All recalls posted in June 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
STERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table
FDA root cause category: Nonconforming Material/Component
What the company is doing
On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.
Recall details
- Recall number
- Z-1941-2010
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- May 14, 2010
- Posted by the FDA
- Jun 30, 2010
- Recalling firm
- Steris Corporation
- Quantity
- 79 units
- Product code
- FQO
- FDA submissions
- K090136
- Distribution
- Worldwide distribution: USA, Australia, India, Korea & New Zealand
Product description
STERIS 4085, 5085 & 5085SRT Surgical Tables
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Model #: 4085, 5085 & 5085SRT Serial Numbers: 0400710041, 0400710042, 0400710043, 0400710044, 0401510077, 0401510080, 0401510081, 0401510082, 0401510086, 0401510087, 0401510088, 0401510089, 0401510090, 0401510092, 0401910133, 0402010001, 0402010002, 0402010003, 0403210095, 0403210096, 0403210096, 0403210097, 0403210098, 0403610046, 0403610047, 0403610048, 0403610049, 0403610050, 0403610053, 0403610054, 0403610055, 0403610057, 0403610061, 0403610062, 0403610063, 0404610061, 0404610062, 0404610063, 0404610063, 0404610064, 0404610064, 0404610065, 0404610065, 0404610070, 0404610070, 0404610071, 0404610071, 0404610072, 0404610072, 0405410037, 0405410038, 0405410039, 0432709078, 0432709079, 0432709079, 0432709080, 0432709081, 0432709082, 0432709083, 0432709083, 0432709084, 0432709084, 0433609001, 0433609001, 0433609002, 0433609002, 0433609005, 0433609005, 0433609062, 0434309100, 0434309101, 0434309103, 0434809001, 0434809003, 0434809004, 0434809005, 0434809006, 0434809007, 0434809007, 0434809008, 0434809008, 0434809012, 0434809013, 0434809014, 0434809014, 0434809015, 0434809015, 0434809016, 0434809016, 0434809017, 0434809017, 0434809018, 0434809018, 0434809019, 0434809019, 0436309084, 0436309085, 0436509122, 0436509123, 0436509124 & 0436509124.
Other devices under the same FDA submission
2 devices cleared under K090136 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- STERIS 5085 SRT Surgical TableSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED1 report
- STERIS 5085 SRT Surgical TableSteris Corporation · TABLE, OPERATING-ROOM, AC-POWERED1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗