STERIS 4085, 5085 & 5085SRT Surgical Tables recallZ-1941-2010 · Steris Corporation · initiated May 14, 2010

Terminated All recalls posted in June 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

STERIS has learned that a limited quantity of manifold assemblies were manufactured by our supplier with incorrect washers and springs in the valves of the manifold. This supplier error has the potential to cause: a) an inability to articulate/move a table section when commanded by the operator; b) two different articulations occurring when a single articulation is commanded, or c) that the table

FDA root cause category: Nonconforming Material/Component

What the company is doing

On May 14, 2010 the firm sent FCL letters to their customers notifying them of the problem. They informed their customers that a service representative will contact them to arrange correction of the unit.

Recall details

Recall number
Z-1941-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 14, 2010
Posted by the FDA
Jun 30, 2010
Recalling firm
Steris Corporation
Quantity
79 units
Product code
FQO
FDA submissions
K090136
Distribution
Worldwide distribution: USA, Australia, India, Korea & New Zealand

Product description

STERIS 4085, 5085 & 5085SRT Surgical Tables

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model #: 4085, 5085 & 5085SRT Serial Numbers: 0400710041, 0400710042, 0400710043, 0400710044, 0401510077, 0401510080, 0401510081, 0401510082, 0401510086, 0401510087, 0401510088, 0401510089, 0401510090, 0401510092, 0401910133, 0402010001, 0402010002, 0402010003, 0403210095, 0403210096, 0403210096, 0403210097, 0403210098, 0403610046, 0403610047, 0403610048, 0403610049, 0403610050, 0403610053, 0403610054, 0403610055, 0403610057, 0403610061, 0403610062, 0403610063, 0404610061, 0404610062, 0404610063, 0404610063, 0404610064, 0404610064, 0404610065, 0404610065, 0404610070, 0404610070, 0404610071, 0404610071, 0404610072, 0404610072, 0405410037, 0405410038, 0405410039, 0432709078, 0432709079, 0432709079, 0432709080, 0432709081, 0432709082, 0432709083, 0432709083, 0432709084, 0432709084, 0433609001, 0433609001, 0433609002, 0433609002, 0433609005, 0433609005, 0433609062, 0434309100, 0434309101, 0434309103, 0434809001, 0434809003, 0434809004, 0434809005, 0434809006, 0434809007, 0434809007, 0434809008, 0434809008, 0434809012, 0434809013, 0434809014, 0434809014, 0434809015, 0434809015, 0434809016, 0434809016, 0434809017, 0434809017, 0434809018, 0434809018, 0434809019, 0434809019, 0436309084, 0436309085, 0436509122, 0436509123, 0436509124 & 0436509124.

Other devices under the same FDA submission

2 devices cleared under K090136 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.