i-STAT PT/INR Cartridge recallZ-1996-2010 · Abbott Point of Care Inc. · initiated May 5, 2010

Terminated All recalls posted in July 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A high mean bias (12-16%) was documented in 8 lots of i-STAT PT/INR cartridges. This may lead a clinician to believe that a patient is adequately anticoagulated when they are not. Inadequate anticoagluation therapy may be given and issues with clotting and/or thrombosis could occur. No reports of injury have been received.

FDA root cause category: Other

What the company is doing

Urgent recall notifications were sent on May 13 2010. Customers were instructed to discontinue use and to return the identified lots. Additional questions are directed to Abbott Point of Care Technical Support at 800-3668020, Option 1 or your Abbott Point of Care representative.

Recall details

Recall number
Z-1996-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
May 5, 2010
Posted by the FDA
Jul 13, 2010
Recalling firm
Abbott Point of Care Inc.
Quantity
44976 in US, 3648 in rest of world.
Product code
GJS
FDA submissions
K020355
Distribution
Worldwide distribution: Argentina, Hong Kong, China, Finland, Greece, India, Portugal, Russian Federation, Saudi Arabia, United Arab Emirates, Italy, Spain, Australia, Germany, UK, Sweden, The Netherlands, France, Canada.

Product description

i-STAT PT/INR Cartridge Abbott Point of Care, Princeton, NJ

Code information (models, lots, serial numbers)

Catalog numbers: 04J50-01, 04J50-02. Lot numbers: N09315, exp 4/28/2010; N09323, exp 5/14/2010; N09323A, exp 5/14/2010; N09346A, exp 5/28/2010; S09347, exp 5/28/2010; S09354, exp 6/14/2010; T10009, exp 6/28/2010; T10011, exp 6/28/2010.

Other devices under the same FDA submission

1 device cleared under K020355 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.