Skip to content

DSD Application Guide (MG04-0001 revision M) for the Pentax… recallZ-2213-2010 · Minntech Corp · initiated Jun 15, 2010

Terminated All recalls posted in August 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Minntech Corp or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The DSD hook up Application Guide incorrectly recommends use of the Medivators DSD-110-HU0109 endoscope hookup to connect the Pentax EG-3630U Ultrasound Endoscope to a Medivators DSD endoscope reprocessor or Medivators Scope Buddy Endoscope Flushing Aid. This may result in inadequate disinfection during reprocessing, and subsequent infection risk to the patient. The recall was due to the misprinte

FDA root cause category: Labeling False and Misleading

What the company is doing

Consignees were sent a Medivators Reprocessing Systems "Urgent Medical Device Recall" letter date June 17, 2010. The letter was addressed to "Dear Medivators Customers". The letter described the problem and product involved. Recommended consignees to immediately examine the Pentax EG-3630U Ultrasound Endoscope hookups and inform all staff of the notice to ensure that the DSD-11--HU01109 is not used with the Pentax endoscope. Also requested consignees to complete and return the enclosed response form.

Recall details

Recall number
Z-2213-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Jun 15, 2010
Posted by the FDA
Aug 12, 2010
Recalling firm
Minntech Corp
Quantity
185
Product code
FEB
FDA submissions
K914145
Distribution
AZ, AR, CA, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MD, MA, MI, MN, MS, MO, NH, NJ, NY, NC, ND, OH, PA, SC, TN, TX, UT, VA, WA, WV, WI, WY.

Product description

DSD Application Guide (MG04-0001 revision M) for the Pentax Universal Video Scope Hookup (DSD-110-HU0109) only when used with the DSD automated endoscope reprocessor (AER) and the Pentax EG-3630U endoscope.

Code information (models, lots, serial numbers)

A lot number listing is not applicable as this is a field correction of the labeling. The 0109 hookup itself is not defective and can still be used safely with other enodscopes.

Other devices under the same FDA submission

3 devices cleared under K914145 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.