CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured… recallZ-2351-2010 · Abbott Laboratories · initiated Apr 16, 2010

Terminated All recalls posted in September 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.

Recall details

Recall number
Z-2351-2010
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 16, 2010
Posted by the FDA
Sep 2, 2010
Recalling firm
Abbott Laboratories
Quantity
69 units total
Product code
GKZ
FDA submissions
K991605
Distribution
Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.

Product description

CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.

Code information (models, lots, serial numbers)

Reservoir Date Codes Dec 17 2009 to March 17, 2010. Part Number/Serial numbers-- CS: 02H30-01/20478AN, 20479AN SL: 02H31-01/22470AK, 21690AK, 22062AK, 21699AK, 20756AK, 22459AK, 21302AK, 21601AK, 22753AK

Other devices under the same FDA submission

3 devices cleared under K991605 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.