Estrone Radioimmunoassay (RIA) Kit, DSL-8700 recallZ-1685-2011 · Beckman Coulter Inc. · initiated Apr 22, 2010

Terminated All recalls posted in March 2011

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Beckman Coulter has confirmed that eight Estrone RIA kit (Ref DSL 8700) lots are inaccurately calibrated to the firm's internal reference calibrator. Lots 992637, 100115A, 100205A, 100205RA, and 100226A have now expired and should not be in active use any more.

FDA root cause category: Other

What the company is doing

Beckman Coulter released an Urgent: Product Corrective Action letter dated July 9, 2010 identifying the affected product and informing consignees of the reason for recall, asking them to take the following actions: *For lots 992637, 100115A, 100205A, 100205RA, and 100226A, review historical external quality control results relevant to the patients' results obtained with these lots to ensure the assay performance was within laboratory specifications. *For lots 100319A, 100409A, 100409RA, review patents' results obtained with the results of external quality control and interpret them in the light of additional applicable tests, patient's clinical status and clinical history of patient. Beckman Coulter has finalized the process of re-adjusting Estrone calibrators' values assignment to the new internal reference calibrator starting with lot 100403A. If consignees need products for additional tests they were told to follow replacement instructions below: *In the United States, please contact Customer Service at 1-800-526-3821 *Outside the United States, contact your local Beckman Coulter Representative. Consignees were asked to share this information with their laboratory staff and retain the notification as part of their laboratory Quality System documentation. They were to also complete and return the enclosed response form within 10 days so the firm can be assured that consignees received the notification. Consignees were instructed to forward any other questions regarding the notification to the DSL Technical Support Center at 1-800-231-7970 in the United States, or contact their local Beckman Coulter Representative.

Recall details

Recall number
Z-1685-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Apr 22, 2010
Posted by the FDA
Mar 17, 2011
Recalling firm
Beckman Coulter Inc.
Quantity
917 units
Product code
CGF
FDA submissions
K935013
Distribution
Worldwide Distribution -- US, including states of CA, FL, GA, IL, IN, KS, MI, NC, NY, OH, OR, TX, VA, WA, and WI and country of Canada.

Product description

Estrone Radioimmunoassay (RIA) Kit, DSL-8700. Intended for in vitro diagnostic use.

Code information (models, lots, serial numbers)

Lot numbers: 992637, 100115A, 100205A, 100205RA, 100226A, 100319A, 100409A, 100409RA

Other devices under the same FDA submission

1 device cleared under K935013 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.