Abbott CELL-DYN 3500 Hematology Analyzer CS (Closed… recallZ-1703-2011 · Abbott Laboratories · initiated Mar 27, 2009
Terminated All recalls posted in March 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
FIRE HAZARD-- Some power supply modules may have incorrect fuses installed, which may result in fires and smoking.
FDA root cause category: Component design/selection
What the company is doing
Abbott Laboratories sent Product Correction letters, dated March 22, 2010, to its customers. The letter listed the additional products that are now included in the recall and gave an explanation of the reason for recall. The letter also discussed the safety impact and provided a list of necessary actions for customers. Customers are to check that the ampere rating of the fuse matches the voltage of operation that is currently being utilized with their CELL-DYN 3000 Analyzer. If the fuse is correct, then the steps provided should be followed. If the fuse is incorrect, then customers are to replace the fuse with the correct fuse and follow the instructions provided. Customers are to contact their local Customer Support Representative in the event that they do not have their replacement fuse that was provided with the Accessory Kit. Customers can refer to the Product Information Letter dated April 20, 2010 for additional labeling to assist with the correct fuse replacement. The Customer Reply form should be completed and the letter should be kept with the Operator's Manual. Questions should be directed to Customer Support at 1-877-4ABBOTT. The firm, Abbott Laboratories, sent a "Product Correction" letter dated March 27, 2009 to its customers. The letter included a Customer Reply form and CELL-DYN accessory kits to inform new customers of the necessity of replacing fuses correctly. The letter described the product, problem and actions to be taken. The customers were instructed to check that the ampere rating of the fuse matches the voltage of operation that is currently being utilized with their CELL-DYN 3700 System and follow the instructions provided, and complete and return the Customer Reply form via fax to 1-800-777-0051 or e-mail [email protected], even if they no longer have the instrument. The firm intends to service items in the field. In April 2010, additional labeling was sent to customers providing additional information when replacing fuses.
Recall details
- Recall number
- Z-1703-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Mar 27, 2009
- Posted by the FDA
- Mar 17, 2011
- Recalling firm
- Abbott Laboratories
- Quantity
- 2961 units
- Product code
- GKZ
- FDA submissions
- K991605
- Distribution
- Worldwide Distribution -- USA, including states of: AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of Argentina, Australia, Bahamas, Barbados, Brazil, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Dominica, Dominican Republic, Ecuador, El Salvador, Germany, Guatemala, Haiti, Honduras, Hong Kong, Jamaica, Japan, Mexico, New Zealand, Panama, Paraguay, Peru, Puerto Rico, Singapore, South Korea, Taiwan, Thailand Trinidad and Tobago, Turks & Caicos, Uruguay, and Venezuela.
Product description
Abbott CELL-DYN 3500 Hematology Analyzer CS (Closed Sampler), CELL-DYN 3500 SL (Sample Loader). Manufactured by Abbott Diagnostics Division, Santa Clara, Ca Medical Device, hematology analyzer for in vitro diagnostic use in clinical Laboratories.
Code information (models, lots, serial numbers)
List numbers: 3500CS: 91340-01, 91340-03; 3500SL: 91350-01, 91350-03. All serial numbers
Other devices under the same FDA submission
3 devices cleared under K991605 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CELL-DYN 02H31-01Abbott Laboratories · Counter, differential cell4 reports
- CELL-DYN 02H31-03Abbott Laboratories · Counter, differential cell3 reports
- CELL-DYN 02H30-01Abbott Laboratories · Counter, differential cell2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 51955).
- Z-1701-2011 · TerminatedAbbott CELL-DYN 3700 Hematology Analyzer CS (Closed Sampler), , and CELL-DYN 3700 SL (Sample Loader), Manufactured by Abbott Diagnostics Division, Santa Clara, Ca Medical Device,…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page