Stratus(R) CS Rotors, catalog number CROTOR, lot number… recallZ-0343-2011 · Siemens Healthcare Diagnostics, Inc. · initiated Dec 7, 2009

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Low frequency defect in the gel filling of this lot that may result in inadequate separation of cells in centrifugation.

FDA root cause category: Process control

What the company is doing

The recalling firm, SIEMENS, issued an "Urgent Field Safety Notice" dated December 2009 to all affected customers. The notice described the product, problem and the action to be taken by the customers. The customers were instructed to review the situation with their laboratory director to determine the need to evaluate previously reported test results; immediately examine their CROTOR inventory and discard all bags that are from lot NH42-251-09 including rotors that were not stored in their original packaging (bag); contact Siemens Customer Service Center at 800-241-0420 to request additional no-charge replacements (Note: A "no-charge" order has already been placed so that one bag of replacement rotors would be recieved shortly); forward this notice to anyone they may have distributed this product, and complete and return the FIELD CORRECTION EFFECTIVENESS CHECK via fax at 302-631-8467. If you have any questions regarding this information, please contact the Siemens Technical Solution Center at 800-405-6473.

Recall details

Recall number
Z-0343-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Dec 7, 2009
Posted by the FDA
Nov 10, 2010
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
330 bags
Product code
JHX
FDA submissions
K984067
Distribution
Worldwide distribution: USA including states of AK, AZ, AR, CA, FL, GA, HI, IA, IL, IN, KY, LA, MI, MN, MO, MS, MT, NC, ND, NH, NM, NY, OH, TX, UT, WA, WI, and WV and country of Canada

Product description

Stratus(R) CS Rotors, catalog number CROTOR, lot number NH42-251-09 Intended use: Stratus(R) rotors is intended for the separation of whole blood samples for analysis on the Stratus(R) CS analyzer

Code information (models, lots, serial numbers)

catalog number CROTOR, lot number NH42-251-09

Other devices under the same FDA submission

1 device cleared under K984067 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.