PROCEED Surgical Mesh recallZ-0903-2011 · Ethicon, Inc. · initiated Oct 18, 2010
Open, Classified All recalls posted in January 2011
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for delamination in one lot of Proceed surgical mesh.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Ethicon sent Urgent Voluntary Product Recall letters via Fed Ex on October 18, 2010. The letters described the issue with the affected product and instructed customers not to use it. They are to immediately remove the recalled product from their inventory and return it according to the instructions provided. A reply card is attached and must be filled out and returned to Stericycle whether or not the customer has product in their possession. Separate correspondence has been sent to end customers stating that if they have purchased the recalled product through a distributor, to return the product directly to Stericycle. Replacement product will be provided at no additional charge for all recalled product.
Recall details
- Recall number
- Z-0903-2011
- Recall class
- Not yet classified by the FDA
- Status
- Open, Classified
- Initiated by the company
- Oct 18, 2010
- Posted by the FDA
- Jan 14, 2011
- Recalling firm
- Ethicon, Inc.
- Quantity
- 87 products
- Product code
- FTL
- FDA submissions
- K060713
- Distribution
- Worldwide Distribution
Product description
PROCEED Surgical Mesh; Product Code PCDG1; :Lot BMG618; Ethicon, Inc. a Johnson & Johnson Company. Used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.
Code information (models, lots, serial numbers)
Product code PCDG1, Lot number BMG618.
Other devices under the same FDA submission
11 devices cleared under K060713 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- PROCEED PCDN1Johnson & Johnson International Inc. · Mesh, surgical, polymeric744 reports
- PROCEED PCDG1Johnson & Johnson International Inc. · Mesh, surgical, polymeric1 recall1 death637 reports
- PROCEED PCDH1Johnson & Johnson International Inc. · Mesh, surgical, polymeric421 reports
- PROCEED PCDM1Johnson & Johnson International Inc. · Mesh, surgical, polymeric154 reports
- PROCEED PCDT1Johnson & Johnson International Inc. · Mesh, surgical, polymeric122 reports
- PROCEED PCDJ1Johnson & Johnson International Inc. · Mesh, surgical, polymeric103 reports
- PROCEED PCDR1Johnson & Johnson International Inc. · Mesh, surgical, polymeric80 reports
- PROCEED PCDB1Johnson & Johnson International Inc. · Mesh, surgical, polymeric54 reports
- PROCEED PCDW1Johnson & Johnson International Inc. · Mesh, surgical, polymeric43 reports
- PROCEED PCDD1Johnson & Johnson International Inc. · Mesh, surgical, polymeric42 reports
- PROCEED PCDL1Johnson & Johnson International Inc. · Mesh, surgical, polymeric30 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page