smith & nephew TRIGEN SURESHOT Distal Targeting System… recallZ-0411-2011 · Smith & Nephew Inc · initiated Oct 15, 2010

Terminated All recalls posted in November 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Targeting software has been found to provide erroneous targeting when used with the 13 mm TAN/FAN Left nails.

FDA root cause category: Other

What the company is doing

Smith & Nephew Inc. sent an URGENT MEDICAL DEVICE CORRECTION letter by FedEx on October 15, 2010, to all customers. The letter was dated October 4, 2010, and identified the product, the problem, and the action to be taken by the customers. Customers were instructed to perform a software upgrade and complete the form provided and return by fax to (901) 566-7835. For questions regarding this recall call (901) 399-5441.

Recall details

Recall number
Z-0411-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Oct 15, 2010
Posted by the FDA
Nov 18, 2010
Recalling firm
Smith & Nephew Inc
Quantity
445 units
Product code
OLO
FDA submissions
K100107
Distribution
Worldwide Distribution - USA including AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, MS, NC, NJ, NM, OH, PA, TN, TX, AND WA and the countries of Australia, Belgium, Canada, China, Denmark, Finland, France, Germany, Italy, Japan, Netherlands, Norway, South Africa, Spain, Sweden, Switzerland, Turkey, and the UK.

Product description

smith & nephew TRIGEN SURESHOT Distal Targeting System, Trauma Interface, V2.0.2, Trauma Launcher V1.01.1 Product is intended to be an intraoperative image-guided localization system. It is a computer-assisted orthopedic surgery tool to aid the surgeon with drill positioning for screws during intramedullary nail implantation.

Code information (models, lots, serial numbers)

All serial numbers prior to and including ACA5

Other devices under the same FDA submission

6 devices cleared under K100107 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.