Skip to content

Medefil Heparin I.V. Flush Syringe 1 units/mL recallZ-0601-2011 · Medefil Incorporated · initiated Nov 1, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medefil Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The heparin lock flush solution was manufactured with contaminated Heparin Sodium USP Active Pharmaceutical Ingredient (API).

FDA root cause category: Material/Component Contamination

What the company is doing

Medefil faxed/emailed the direct accounts on 11/1/10, and sent the accounts a hard copy of the Urgent Product Recall letter, dated 11/1/2010, on 11/2/10. The accounts were informed of Medefil's recall of 18 lots of Heparin I.V. Flush Syringes produced from the active pharmaceutical ingredient (API) Heparin Sodium, USP, lot 1035-0780, that has been recalled by the manufacturer, Scientific Protein Laboratories, due to the presence of a contaminant, Oversulfated Chondroitin Sulfate. The accounts were instructed to stop use of the 18 lots of Heparin I.V. Flush Syringes immediately, place them under quarantine, and contact Medefil to arrange for the product to be returned. The accounts were requested to complete and return to Medefil the enclosed response form, indicating the quantity on hand of each recalled lot. Any questions were directed to Pradeep Aggarwal, President at 630-682-4600.

Recall details

Recall number
Z-0601-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 1, 2010
Posted by the FDA
Dec 10, 2010
Recalling firm
Medefil Incorporated
Quantity
339,180 syringes
Product code
NZW
FDA submissions
K092491
Distribution
Nationwide Distribution --- including states of Arizona, California, Florida, Illinois, Ohio, Louisiana, Massachusetts, Michigan, Minnesota, New York, Texas, and Virginia.

Product description

Medefil Heparin I.V. Flush Syringe 1 units/mL; a polypropylene luer locking syringe containing Heparin Lock Flush Solution, USP, a sterile non-pyrogenic, isotonic injection; each mL in a Heparin I.V. Flush Syringe contains 1 units of heparin sodium derived from porcine intestinal mucosa and 9 mg of sodium chloride in water for injection; Medefil, Inc. Glendale Heights, IL 60139. The pre-filled syringes were packed as 60 syringes per dispenser box, 16 boxes per shipping carton in the following syringe configurations: a) NDC# 64253-444-22, catalog # MIH-4422: 2 mL fill in 6 mL syringe; b) NDC# 64253-444-25, catalog # MIH-4425: 5 mL fill in 6 mL syringe

Code information (models, lots, serial numbers)

a) catalog # MIH-4422: Lot H210209, exp. date 04/12; Lot H210238, exp. date 05/12; Lot H210265, exp. date 05/12; b) catalog # MIH-4425: Lot H210210, exp. date 04/12; H210241, exp. date 05/12.

Other devices under the same FDA submission

3 devices cleared under K092491 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 57096).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page