Care E Vac 3 Portable Suction recallZ-0833-2011 · Ohio Medical Corporation · initiated Nov 8, 2010

Terminated All recalls posted in December 2010

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ohio Medical Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The D-14 diode on the circuit board may fail, resulting in a damaged, overheated circuit board. Also a screw used to attach one of the feet of the unit may damage the battery inside the device if the screw is over-tightened.

FDA root cause category: Device Design

What the company is doing

Ohio Medical sent Urgent Medical Device Correction Notification letters dated November 4, 2010, to all affected customers on November 8, 2010. The letters identified the product, the problems, and the action to be taken by the customer. Customers were instructed to take the affected product out of service, remove diode 14, and replace the foot screw contained therein. Customers were instructed to visit the micro-sites established for these repair procedures for step-by-step instructions on performing the repairs. Accounts with questions after viewing the video demonstration were directed to the Ohio Medical technical assistance line at 224-430-1522. Customers were asked to complete the attached Customer Acknowledgment Form and return the form to the Quality Assurance Department at (fax) 1.847.855.6304, or e-mail to: [email protected], or mail to: Ohio Medical Corproation Quality Assurance 1111 Lakeside Drive Gurnee, Illinois 60031. For questions regarding this recall call 847-855-6270.

Recall details

Recall number
Z-0833-2011
Recall class
Not yet classified by the FDA
Status
Terminated
Initiated by the company
Nov 8, 2010
Posted by the FDA
Dec 29, 2010
Recalling firm
Ohio Medical Corporation
Quantity
4,183 units
Product code
BTA
FDA submissions
K062610
Distribution
Worldwide Distribution - USA nationwide, Puerto Rico, and the coutries of Canada, Israel, Jordan, Kenya, Kuwait, Saudi Arabia, Taiwan, Thailand and Vietnam.

Product description

Care E Vac 3 Portable Suction; Battery/AC Portable Aspirator; Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, IL 60031; Model 758000 The device is a portable aspirator to help evacuate saliva, mucous, vomit or other aspirant from the mouth and or airway to allow adequate respiration or ventilation of the patient.

Code information (models, lots, serial numbers)

Model 758000; Diode only: serial numbers CDF00000001 thru CDG00001052;  Screw only: serial numbers CDJ00000745 thru CDK00000667 Diode & replacement screw: serial numbers CDG00001053 thru CDJ00000744

Other devices under the same FDA submission

1 device cleared under K062610 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.