CyberKnife Robotic Radiosurgery System, medical… recallZ-0714-2011 · Accuray Inc · initiated Oct 28, 2010
Terminated All recalls posted in December 2010
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Accuray Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If users modify the dose calculation box to cover just the contoured anatomy and not the entire CT volume, dosage may not be properly calculated.
FDA root cause category: Other
What the company is doing
Accuray Inc., sent an "URGENT DEVICE CORRECTION" letter dated October 21, 2010, to all customers. The letter described the product, problem and actions to be taken by the customers. The customers were instructed to increase the size of the dose box to include CT volume, not just the contoured anatomy; recalculate dose at high resolution, and examine the treatment plan to ensure that applicable organs at risk are accounted for and all portions of the patient receive the desired level of doses. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at +1-877-668-8667 (USA), +1-408-716-4700 (non-USA) or email: [email protected].
Recall details
- Recall number
- Z-0714-2011
- Recall class
- Not yet classified by the FDA
- Status
- Terminated
- Initiated by the company
- Oct 28, 2010
- Posted by the FDA
- Dec 16, 2010
- Recalling firm
- Accuray Inc
- Quantity
- 219 units
- Product code
- IYE
- FDA submissions
- K091999
- Distribution
- USA and countries of: Athens, Belgium, Canada, China, Czech Republic, France, Germany, India, Italy, Japan, Korea, Malaysia, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Thailand, The Netherland, Turkey, Ukraine, United Kingdom, Vietnam, Venezuela.
Product description
CyberKnife Robotic Radiosurgery System, medical charged-particle radiation therapy system, manufactured by Accuray Inc., Sunnyvale, CA. Indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Code information (models, lots, serial numbers)
All versions
Other devices under the same FDA submission
1 device cleared under K091999 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CyberKnife® VSI™ System 032500-M-000Accuray Incorporated · Accelerator, Linear, Medical1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.